
© Isothermal Technology
Page 24 of 102
923 milliK manual - issue: 1.10
3.2.1 Configuring milliK to Measure Temperature with a Thermistor
Firstly the thermistor and its calibration details must be entered into the milliK’s thermometer
database (see section 3.9.1), the thermistor will then appear in the ‘Thermometer’ list. Press
the
Settings
(F3) key to access the
Settings
Window, which opens with the ‘Channel’
parameter selected. Use the
←
←
←
←
→
→
→
→
keys to select the channel that you want to configure. Then
use the
↑
↑
↑
↑
↓
↓
↓
↓
keys to select each parameter in turn and set it to the required value either by
using the
←
←
←
←
→
→
→
→
keys to sequence through the options or pressing the
↵
↵
↵
↵
key to open a new
Window with all the options listed. A typical configuration for a thermistor would be:
The Samples/Reading parameter is the number of samples averaged together to form a single
reading. Increasing this value will reduce the noise on the measurements (standard deviation
will reduce as the square root of the number of samples) but also slow down the update rate.
The resistance of a thermistor can be measured without entering any calibration information
by declaring the ‘Thermometer’ as a “Default Thermistor”. This can be useful when
calibrating thermistors against other thermometers: