17
Ref
#
Command
Argument
Function
Op
Response
The
pump
display
is
under
the
control
of
the
D
and
DA
commands
and
keypad
entry
at
the
pump
is
disabled.
8
D
‐‐‐‐‐
(5)
Numeric
digits
to
write
to
the
control
panel.
Write
numbers
to
the
pump
to
display
while
under
external
control.
Set
*
9
DA
‐‐‐‐
(4)
Alphabetic
characters
to
write
to
the
control
panel.
Write
letters
to
the
pump
to
display
while
under
external
control.
Set
*
10
L
Set
pump
to
RPM
mode.
Set
*
11
M
Set
pump
to
Flow
Rate
mode.
Set
*
12
N
Set
pump
to
Time
mode.
Set
*
13
O
Set
pump
to
Volume
(at
rate)
mode.
Set
*
14
]
Set
pump
to
“pause”
mode.
This
essentially
stops
the
pump.
Set
*
15
P
Set
pump
to
Time
+
Pause
mode.
Set
*
16
Q
Set
pump
to
Volume
+
Pause
mode.
Set
*
17
G
Set
pump
to
Volume
(over
time)
mode.
Set
*
‐
volume
too
small
and
time
too
long
+
volume
too
large
and
time
too
short
Parameter
Control
18
E
Returns
whether
the
pump
is
currently
running
or
not.
Get
+
running
‐
stopped
19
#
Returns
the
following
fields,
each
separated
by
a
space:
Pump
model
description
(variable
length)–
A
text
field
describing
the
model
of
pump.
This
description
may
contain
spaces.
Pump
software
version
(3
digits)
–
The
version
of
software
currently
running
in
the
pump.
Pump
head
model
type
code
(3
digits)
–
A
code
describing
the
type
of
pump
head
installed.
The
first
digit
represents
the
number
of
channels
for
the
pump,
and
the
second
2
digits
represent
the
number
of
rollers.
Get
[model][SP][s/w
ver][SP][head
code][CR][LF]
20
(
Returns
the
pump
software
version
–
The
version
of
software
currently
running
in
the
pump.
This
value
is
returned
in
4
characters
with
the
value
right
justified
and
zero
‐
padded.
Get
0123[CR][LF]
21
)
Returns
the
pump
head
model
type
code
–
A
3
digit
code
describing
the
type
of
pump
head
installed.
The
first
digit
represents
the
number
of
channels
for
Get
0123[CR][LF]