10
6.4.5
GLOBAL
SETTINGS
Menu
In
the
GLOBAL
SETTINGS
Menu
the
display
language
can
be
selected
between
English,
French,
Spanish,
and
German.
The
pump
can
also
be
reset
to
the
factory
defaults
at
this
menu.
6.5
Installing
Tubing
1.
Switch
the
pump
off
2.
Remove
the
cassette
by
slightly
pressing
the
fixing
‐
tongue
and
lifting
it
simultaneously
3.
Insert
the
2
‐
stop
or
3
‐
stop
color
‐
coded
tubing
with
one
stopper
into
the
cassette
4.
Let
the
tubing
hang
down
(prevent
it
from
being
twisted)
5.
Insert
the
tubing
with
the
second
stopper
at
the
other
end
of
the
cassette
6.
Reinsert
the
cassette
into
the
roller
‐
head
7.
Make
sure
each
cassette
is
positioned
and
aligned
well
between
dedicated
segments
on
the
bars
8.
Turn
the
pump
on
9.
When
the
pump
is
idle,
release
all
cassettes
to
maximize
tubing
life
(beware
of
siphoning
or
back
flow)
6.6
Information
Screens
Indicates
that
the
pump
is
being
remotely
controlled
by
a
PC
via
USB
connection.
Indicates
that
the
internal
temperature
has
exceeded
a
safe
threshold
and
pumping
will
stop
automatically
to
protect
the
pump.
Turn
the
pump
power
switch
off
and
check
your
parameters,
differential
pressure
and
pump
head
for
blockage.
The
pump
will
not
operate
in
an
overheated
condition.
Allow
the
pump
to
cool
before
turning
the
power
switch
back
on.
6.7
Differential
Pressure
The
Reglo
ICC
can
be
used
for
continuous
duty
at
a
maximum
differential
pressure
of
1.0
bar
(smaller
tube
sizes
and/or
lever
type
cassettes
allow
higher
pressures).
he
pump
is