9
6.4.4
CALIBRATION
Menu
In
the
CALIBRATION
Menu
each
channel
can
be
calibrated
independently.
The
channel
being
calibrated
is
listed
in
the
upper
right
hand
corner.
The
Reglo
ICC
allows
each
channel
to
be
set
up
and
calibrated.
This
allows
for
maximum
flexibility
and
optimum
channel
to
channel
precision.
Prior
to
the
calibration
each
channel
must
be
configured
for
tubing
ID,
the
number
of
backsteps
that
are
desired
for
drip
free
dispensing,
and
calibration
of
that
channel.
The
Reglo
ICC
allows
you
to
enter
the
volume
of
fluid
to
dispense
for
calibration,
dispense
time
for
the
specified
volume,
and
pumping
direction.
Once
you
have
entered
those
values,
dispense
desired
volume
into
a
container.
Measure
the
actual
amount
of
the
fluid
by
weight
or
volume
and
enter
that
value
on
the
last
screen
to
adjust
the
pump
for
your
application.
The
calibration
can
be
interrupted
by
pressing
Run/Stop
button
which
will
cancel
the
existing
progress.
New
uninterrupted
calibration
run
will
be
required
in
order
to
calibrate
the
channel
properly.
When
adjusting
the
dispense
time
for
calibration,
if
requested
time
is
shorter
than
the
pump’s
physical
capabilities
the
dispense
time
will
automatically
default
to
minimum
achievable
dispense
time.
You
can
complete
the
calibration
by
pressing
Ok/Enter
button
following
your
measurement
input.
You
will
receive
a
message
and
checkmark
once
your
calibration
is
successful.
You
can
cancel
the
calibration
after
the
calibration
run
by
pressing
the
Reset
button.
You
will
receive
the
message
displaying
that
the
calibration
has
been
canceled.