background image

ISKO KOCH GmbH, Egerländer Str. 28, 95448 Bayreuth 

 

Instruction Manual page  14 to 21 

ISKO KOCH GmbH, Egerländer Str. 28, 95448 Bayreuth 

7

 

Starting the Standing Frame 

Connect the power cable to a suitable supply and switch on the socket, the Frame is ready for use.  

The additional forced switch-off is released, as long as the emergency stop push button, pos. 1 (see 
chapter 4.2), is kept pressed. The Frame can be operated again, only after this emergency stop push 
button has been released. Up- or downward movement of the Lift Chord is effected by pressing the 
respective key of the manual control switch, pos. 16 (see chapter 4.2). The keys are marked with arrows 
according to their functions (see chapter 4.2) 

 

 

How to stand up using "DELTA 2004“

 

Set  up the Standing Frame and fit the  lifting harness as described earlier. Plug the  lifting 
mechanism into the electricity supply and check the unit is functioning by checking that the 
green indicator light on the front of the unit is on. Press the 'down' arrow on the handset to 

feed out the Lift Chord and connect them to the lifting harness. Press the ‘up’ button to 

remove the slack from the Lift Chords before proceeding to lift.  

The  'down'  button  on  the  handset  will  override  the  'up'  button.  If  the  user  or  assistant 
operating the handset wants the lifting mechanism to lower the user back to sitting, pressing 
the 'down' button will lower the user, even if the 'up' button is still depressed. 

The handset connection to the motor unit is on the side of the box on the front of the motor unit. If 
the handset ever gets detached from the motor unit, it can be re-connected to the socket on the side 
of the unit. If the handset is attached to either of the sockets on the front of the unit it will not work 
and may cause damage to the motor. 

If  the  user  has  sufficient  arm  and  trunk  control  they  can  guide  the  lift  to  standing  themselves.  If 
additional  assistance  is  required,  an  appropriately  trained  person  can  assist  the  lift  to  standing.  A 
reassuring hold can be placed around the user's trunk to help guide them to vertical. In either case, 
pressing the 'up' arrow on the handset takes up the slack in the Lift Chords and lifts the user's pelvis as 
the knees are blocked against the knee supports. Once in a vertical position, the button can be released 
to hold the position. A separate hip support unit can be fitted in position to provide more positive hip 
control.  

Always leave the lifting straps and harness in position when the user is standing. 

 

To return the user to sitting, provide an appropriate level of support and press the 'down' button on 
the handset. Guide them to a sitting position and release the button when the sitting surface is fully 

Pic. 7.1

 

Содержание Delta 2004

Страница 1: ...mbH Egerländer Str 28 95448 Bayreuth Owner s Manual KR 204 B S Standing Frame with electric retractable belt Modell Delta 2004 KR 214 B S Standing Frame with electric retractable belt and tilting function Model Delta Evolution Stand 10 2021 Rev 1 4 KR 204 S Delta 2004 Delta Evolution KR 214 S ...

Страница 2: ...se of this Standing Frame we ask you and your assistant to carefully read through the owner s manual and keep it easily accessible EU Declaration of Conformity We as solely responsible manufacturers declare that our Standing Frame meets the basic standards of the EU guideline for Medical Product 2017 745 Appendix II ISKO KOCH GmbH Egerländer Str 28 95448 Bayreuth Please read completely and careful...

Страница 3: ... Instructions for Adjustment 9 Body Height 9 Back Pad 10 Foot and Heel Strap 10 Depth Adjustable Footboard 11 Sliding Tray 12 Knee Pads 12 Lifting Harness 13 6 Approaching the Standing Frame 13 7 Starting the Standing Frame 14 How to stand up using DELTA 2004 14 How to stand up using DELTA Evolution 15 What to do when the electrical supply is cut off 17 Suck blow function for using the standing fr...

Страница 4: ...H GmbH Egerländer Str 28 95448 Bayreuth Instruction Manual page 4 to 21 ISKO KOCH GmbH Egerländer Str 28 95448 Bayreuth 14 Recommended Accessories 19 15 Maintenance Intervals 20 16 Instructions for Disposal 21 ...

Страница 5: ...the symbol together with the name and address of the manufacturer i e the person who places the medical device on the market Conformity symbol according to 2017 745 of the Medical Device Directive Shows the medical device provided by the manufacturer in accordance with EU Directives 2017 745 Device type B according to IEC 601 1 Special protection against electric shock Device of protection class I...

Страница 6: ...air arms of the hip bar rail will lead to damage 3 General product description The standing frame is used exclusively for standing training for people who are unable to walk or have difficulty walking for independent use with an assistant Standing training is used in inpatient and outpatient rehabilitation as well as in the home environment The construction of our standing devices enables independ...

Страница 7: ...ice please make sure that the voltage and frequency of your mains supply correspond to the specifications on the type plate Make sure that the floor is level when choosing a location for the standing frame Provide a suitable floor covering if the standing frame has to be moved frequently Carpets rugs and loose floor coverings can be damaged or make it difficult to push the appliance Connect the po...

Страница 8: ...ISKO KOCH GmbH Egerländer Str 28 95448 Bayreuth Instruction Manual page 8 to 21 ISKO KOCH GmbH Egerländer Str 28 95448 Bayreuth Description of the individual parts of the Standing Frame ...

Страница 9: ...t the moulded crescent pad of the table top is located between sternum and upper edge of the pelvis in accordance with the individual personal requirements By no means it is allowed to adjust the pad of the table top so as to be located below the upper edge of the pelvis Body Height The height adjustment of the table pos 10 pic 5 1 and the handle bars pos 3 pic 5 1 are used for adjusting the table...

Страница 10: ...g handles Ensure the rails are firmly positioned before using The lifting harness will hold the user in a standing position However additional hip supports can be used to provide more positive pelvic control The hip support bar is positioned on the hip support rails which allows adjustment of the depth of the hip pads in relation to front of the tray The hip support bar is semi permanently attache...

Страница 11: ...itted which allows for greater flexibility of foot positioning This option may be required by those with fixed flexion deformities at the knees and hips or for some designs of wheelchair foot supports The above description of setting up the standing frame still applies However the footboard can be slid forwards and backwards to improve foot positioning by undoing the knob pos 8 pic 5 6 in the midd...

Страница 12: ...tion With this in mind adjust the tray forward if necessary to accommodate their position If the tray is adjusted correctly the user will touch the safety cut out button once they are in a comfortable upright position Knee Pads The knee pad s height can be adjusted by loosening the vertical latch pos 14 chapter 4 2 and its depth by loosening the horizontal latch pos 15 chapter 4 2 Knee support sho...

Страница 13: ...orizontal in the sitting position with feet on the footboard The Delta should be positioned as close to the user as possible The cut outs in the footboard are designed to receive the front wheels of a wheelchair Lock the castors on the Delta before using the DELTA User s feet should be positioned on the footboard with the velcro toe straps used to help maintain position and their knees should be f...

Страница 14: ...istant operating the handset wants the lifting mechanism to lower the user back to sitting pressing the down button will lower the user even if the up button is still depressed The handset connection to the motor unit is on the side of the box on the front of the motor unit If the handset ever gets detached from the motor unit it can be re connected to the socket on the side of the unit If the han...

Страница 15: ...traps from the harness by depressing the red button on each buckle and pulling the two items apart How to stand up using DELTA Evolution The Standing Frame with tilting function facilitates easier access and a smoother lifting operation Arm rails knee support and footboard are inclined towards the patient prior to standing up see Pic 7 2 Pic 7 2 Pic 7 3 Pic 7 4 Pic 7 5 ...

Страница 16: ...ghtened Please note the user has not yet begun to lift 3 By pressing the standing up switch the two motors are actuated simultaneously and the standing up process is started see pics 7 2 7 5 Sitting down process Description Key on the handset 1 Once you press the sitting switch as indicated left the sitting down process proceeds The twodrive units move again simultaneously and the patient is broug...

Страница 17: ...ight pressure on the user s body The assistant s left hand reaches under the user s buttocks to support the user when putting the user back into the wheelchair Suck blow function for using the standing frame The electrical components can also be operated with a suck blow function When suck blow function is applied to the mouthpiece of the tube the straps are moved upwards and downwards when blowin...

Страница 18: ...100 125 cm 100 125 cm 100 125 cm Design description The standing frame is manufactured as a welded steel tube construction The surfaces are powder coated or galvanized All wooden parts are either laminated or painted 10 Maintenance and Care For cleaning the pipe parts and the wooden parts with a damp cloth all household cleaners without ammonium chloride and scouring agents are permissible Solvent...

Страница 19: ...exceed or fall below the concentrations indicated They must not contain corrosive or caustic components They must not contain substances that alter the surface structure or the adhesion properties of the materials Lubricants must not be attacked by the cleaning and disinfecting agent Under no circumstances should soap or washing active substances be added to the disinfectant In the case of product...

Страница 20: ... X Lift support harness Size XL with 10 cm lip KR 609 2 X Lift support harness Size XL KR 609 3 X 15 Maintenance Intervals After having had the Standing Frame in operation for two years at the latest a thorough visual inspection a functional test and a current leakage test according to DIN EN 62535 should be effected In particular please consider the following issues when effecting the visual insp...

Страница 21: ...GmbH Egerländer Str 28 95448 Bayreuth 16 Instructions for Disposal Disposal of the device and its accessories if any may only be carried out in accordance with waste segregation regulations Please observe that all electrical components should be disposed of in accordance with legal guidelines ...