5 NHR, SHR, NHS and MICC
This manual uses the notation NHR or NHS if there is a special option for NHR or NHS and uses the notation NHR/S for common
things. Commands that apply to MICC only are marked accordingly too.
5.1. SCPI Instruction Set
For the list of common commands, see section 4: Common instruction set.
Command, for channels
Description
:
VOLT
age
˽
<Voltage>[V],(@<Channel>)
Set the channel voltage set V
set
in Volt
MICC: If the channel is configured with EPU, the voltage sign
defines the polarity of the output voltage.
Example: :VOLT 1234,(@0)
˽
˽
ON
,(@<Channel>)
Switch on High Voltage with the configured ramp speed
Example: :VOLT
˽
ON,(@1)
˽
OFF
,(@<Channel>)
Switch off High Voltage with the configured ramp speed
Example: :VOLT
˽
OFF,(@1)
˽
EMCY
˽
OFF
,(@<Channel>)
Shut down the channel High Voltage (without ramp). The channel
stays in Emergency Off until the command EMCY CLR is given.
˽
Example: :VOLT EMCY OFF,(@2)
˽
˽
˽
EMCY
˽
CLR
,(@<Channel>)
Clear the channel from state emergency off. The channel goes to
state off.
Example: :VOLT EMCY CLR,(@2)
˽
˽
:
BOU
nds
˽
[V],(@<Channel>)
Set the channel voltage bounds V
bounds
in Volt
Example: :VOLT:BOUNDS 10,(@3)
˽
:CURR
ent
˽
<Current>[A], (@
˽
<Channel>)
Set the channel current set I
set
in Ampere
Example: :CURR 1E-3,(@1)
˽
:BOU
nds
˽
<Current>[A],(@<Channel>)
Set the channel current bounds I
bounds
in Ampere
Example: :CURR:BOUNDS 100E-6,(@3)
˽
:EV
ent
˽
CLEAR
,(@<Channel>)
Clear the Channel Event Status register (8.3).
Example: :EVENT CLEAR,(@4)
˽
:MASK
˽
<Word>,(@<Channel>)
Set the Channel Event Mask register (8.4).
Example: :EVENT:MASK 65535,(@4)
˽
SCPI protocol common instruction set | Last changed on: 2020-11-10 |
www.iseg-hv.com
25/63