8.3. Channel Event Status
Access:
read-write
Data-Type:
UI4
Command:
:READ:CHAN
nel
:Ev
ent
:STAT
us
?
:EV
ent
˽
CLEAR
Bit31
Bit30
Bit29
Bit28
Bit27
Bit26
Bit25
Bit24
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Reserved
Bit23
Bit22
(1
Bit21
(1
Bit20
(1
Bit19
(1
Bit18
(1
Bit17
(1
Bit16
(1
Reserved
Event Voltage
Bound Lower
Event Voltage
Bound Upper
Event Voltage
Ramp Down
Event Voltage
Ramp Up
Event Current
Ramp Down
Event Current
Ramp Up
Event End of
Current Ramp
Bit15
Bit14
Bit13
Bit12
Bit11
Bit10
Bit09
Bit08
Event Voltage
Limit
Event Current
Limit
Event Current
Trip
Event External
Inhibit
Event Voltage
Bounds
Event Current
Bounds
Event Arc
Error
Reserved
Bit07
Bit06
Bit05
Bit04
Bit03
Bit02
Bit01
Bit00
Event
Constant
Voltage
Event
Constant
Current
Event
Emergency
Off
Event End Of
Voltage Ramp
Event On To
Off
Event Input
Error
Event Arc
Reserved
Notes:
1)
NHR and SHR only
The Channel Event Status contains the captured status. An event bit is set if the corresponding Channel Status bit is 1 or is
changing to 1.
Different to the status bit an event bit isn't automatically reset. A reset has to be done by the user by writing an 1 to this event bit.
All channel events can be cleared by the
command. With the command *CLS all module and channel event
registers are cleared at once.
Some event bits also perform a blocking function, which means if this Bit and it’s corresponding event mask bit is set, then:
•
The voltage set value can not be increased while the high voltage is on
•
After the high voltage is turned off, it cannot be turned on again
After either the event bit or the corresponding mask is cleared, the blocking is disabled again.