background image

Doc. SIE20167   

 

 

 

 

Rev. 2   

 

 

Page 

6 of  11 

2 APPLICABLE STANDARDS 

The  test  set  conforms  to  the  EEC  directives  regarding  Electromagnetic  Compatibility  and  Low 
Voltage instruments. 

2.1 Electromagnetic Compatibility 

Directive  no.  2004/108/EC. Applicable Standard : EN61326-1 + A1 + A2. 

EMISSION 

- EN61000-3-2: harmonic content induced into the power supply: class A. 
- EN 61000-3-3: Limitation of voltage fluctuations and flicker. Acceptable limits: basic. 
- CISPR16 (EN 55011 class A): Limits and measurement methods of radio-electric disturbances 
for industrial, medical and scientific instruments at radio-electric frequencies. 
Acceptable limits for conducted emission: 
. 0.15-0.5 MHz:  

79 dB pk; 66 dB avg. 

. 0.5-5 MHz:       

73 dB pk; 60 dB avg. 

. 5-30 MHz:   

73 dB pk; 60 dB avg. 

Acceptable limits for radiated emission: 
. 30-230 MHz:  

40 dB (30 m) 

. 230-1000 MHz:   

47 dB (30 m) 

IMMUNITY 

- EN 61000-4-2: Immunity tests for ESD. Test values: 8 kV in air; 4 kV in contact. 
- EN 61000-4-3; Immunity tests for radio frequency interference. Test values (f= 900 

 5 MHz): 

field 10 V/m, modulated AM 80%; 1 kHz 
- EN 61000-4-4; Immunity tests for high speed transients (burst). Test values: 2 kV peak; 5/50 ns. 
-  EN 61000-4-5;  Immunity tests  for surge. Test  values: 1 kV peak differential  mode; 2 kV peak 
common mode; 1.2/50 us. 
-  EN  61000-4-6:  immunity  to  low-voltage  sinusoidal  waveform.  Test  values:  0.15-80  MHz,  3 
Vrms, 80% AM 1 kHz. 
- EN 61000-4-8: Immunity tests for low frequency magnetic fields.  Test values: 30 Arms/m. 
- EN 61000-4-11: Immunity test for power supply drops. Test value: 1 cycle; 100% drop. 

2.2 Low voltage directive 

-  Directive n.  2006/95/EC. 
Applicable standards, for a class I instrument, pollution degree 2, Installation category II:  CEI EN 
61010-1. In particular: 
- Dielectric Rigidity:  1.4 kV, 1 minute. 
- Isolation resistance: > 100 Mohm @ 500 V DC. 
- Earth resistance : < 0.1 Ohm. 
- Dispersion current: < 5 mA. 
- Inputs/outputs protection: IP 20  - IEC 60529. 
- Acoustic noise: < 75 dB, at full power. 
- Operating temperature: 0 - 40°C; storage: -25°C to 70°C. 
- Relative humidity : 5 - 95%, not condensing. 
- Vibration: IEC 68-2-6 (20 m/s^2 at 10 – 150 Hz); 
- Shock: IEC 68-2-27 (15 g; 11 ms; half-sine). 
- Altitude: less than 2000 m. 

 

Shop for Power Metering products online at:

1.877.766.5412

www.

PowerMeterStore

.ca

Содержание BTS200 MKII

Страница 1: ...Doc SIE20167 Rev 2 Date 27 03 2013 BTS200 MKII BATTERY TEST SET Shop for Power Metering products online at 1 877 766 5412 www PowerMeterStore ca...

Страница 2: ..._____________________ 6 2 2 Low voltage directive _______________________________________________________________________ 6 3 PERFORMANCE SUMMARY_______________________________________________________...

Страница 3: ...he external shunt it is also possible to use the real burden as the additional burden and to control the total discharging current in the selected mode The device is microprocessor based This allows t...

Страница 4: ...a shunt If a test is interrupted for instance because of a faulty contact or cell after the repair it is possible to continue the test this was not possible with the former device The test set firmwar...

Страница 5: ...ues f 900 5 MHz field 10 V m modulated AM 80 1 kHz EN 61000 4 4 Immunity tests for high speed transients burst Test values 2 kV peak 5 50 ns EN 61000 4 5 Immunity tests for surge Test values 1 kV peak...

Страница 6: ...le encountered during the day If during the discharge any of the programmed limits is trespassed the test is immediately stopped and the alarm contact closed Once the test is finished it is possible t...

Страница 7: ...limits 84 to 140 V DC 240 V range limits 140 to 270 V DC Maximum starting voltages on the 24 V range 20 to 35 V DC Maximum starting voltages on the 48 V range 34 to 65 V DC Maximum starting voltages o...

Страница 8: ...sors Thermal protection Independent from the microprocessors Dimensions 283 x 700 x 420 mm WDH Weight 39 kg without transport case Transport wheels YES Transport case dimension 330 x 900 x 600 mm WDH...

Страница 9: ...External discharge current Battery voltage Ah discharged Elapsed test time Test stop settings Minimum voltage Maximum Ah Maximum test time System alarms Test stop from remote contact Continue after a...

Страница 10: ...nges 40 A DC 10 mV A and 400 A DC 1 mV A Accuracy 1 5 up to 40 A 2 up to 400 A Maximum conductor diameter 30 mm PII12167 ELU200 MKII Same as BTS200 MKII without the local control PII21167 The followin...