Operation
13103F-EN
23
Using the Two-Position Eye Safety Filter
Closing the filter
When the filter is closed, the filter moves into your visual path and your
eyes are protected from laser radiation above regulatory requirements.
Before beginning treatment, close the filter using the lever on the
two-position eye safety filter.
EndoProbe
displays on the OcuLight GL/GLx
console status panel when the filter is closed;
ENT Probe
displays on the
OcuLight OR.
Opening the filter
When the two-position eye safety filter is open, the filter moves out of your
visual path and you cannot activate the laser.
Green Safety Filter?
displays
on the OcuLight GL/GLx and
Two-Position Safety Filter?
on OcuLight OR
consoles status panel when the filter is open and the yellow Standby light
flashes.
16
6
Lever
Open
Closed