EndoProbe and ENT Probe Operator Manual
13103F-EN
14
Installing a two-position eye safety filter
OcuLight GL/GLx console
The two-position eye safety filter is compatible with firmware versions
1.05, 1.1, and higher. When installing this type of filter, first configure your
console:
1. Activate the diagnostic mode to determine which firmware version you
have by pressing the Counter Reset and Treat/Standby buttons
simultaneously for three seconds or until you hear a beep. The Treat
and Standby lights illuminate.
2. Simultaneously press and hold the Counter Reset button while rotating
the Repeat Interval control clockwise until you see
#7
in the Counter
display. Release the Counter Reset button. Your version of firmware is
shown in the status display.
3. If you have a compatible version of firmware, turn the Repeat Interval
control until
#8
is shown in the Counter display. Press the Treat/
Standby button to scroll to
Movable.
4. Press the Mode button to activate your selection and exit diagnostic
mode.
OcuLight OR
1. Refer to “Setting Up for an Endo Procedure” in the OcuLight OR
Operator Manual.
OcuLight TX
1. Simply connect the SmartKey to the laser.. No changes to laser
preference setting required. The OcuLight TX is designed to recognize
a twp-position filter.
WARNING
The two-position eye
safety filter is not
compatible with
OcuLight GL/GLx firmware
versions 1.02, 1.03, 1.04, or
1.06.
Содержание EndoProbe
Страница 1: ...EndoProbe and OtoProbe Operator Manual IRIDEX 13103 EN Revision F...
Страница 6: ...iv EndoProbe and OtoProbe Operator Manual 13103F EN...
Страница 10: ...4 EndoProbe and ENT Probe Operator Manual 13103F EN...
Страница 43: ...37 Clinical Safety 13103F EN Sample of an ENT sterile pouch label...
Страница 44: ...38 EndoProbe and ENT Probe Operator Manual 13103F EN...