34
EndoProbe and ENT Probe Operator Manual
13103F-EN
Ensuring ocular protection
Protection for the physician
To protect your eyes from laser radiation when using an operating
microscope with the probe, install the appropriate eye safety filter(s).
Protection for all persons in the treatment room — laser safety
eyewear requirements
When using the console with a probe, a Laser Safety Officer should
determine the need for safety eyewear for others in the treatment room
based on the MPE, Nominal Ocular Hazard Area (NOHA), and Nominal
Ocular Hazard Distance (NOHD) for the probe and the console used and
the configuration of the treatment room.
WARNING
Ensure that all persons in
the treatment room are
wearing the appropriate
laser safety eyewear. Never
substitute prescription
eyewear for appropriate
laser safety eyewear.
REFERENCE
See your console manual
for more information
about the formula used to
calculate the worst case
NOHD for the console in
use.
REFERENCE
For further information,
you may refer to:
IEC 60825-1 and
ANSI Z136.1.