EN
24
Monitor Display
Mode
INR
or
INR + PT
Example Low
QC1 Display
Example High
QC1 Display
A “LO” or “HI” QC message appears when the test strips have expired, or have not been stored
properly, and have deteriorated. If you get a “LO” or “HI” QC message, check to make sure your
test strips have not expired, that the correct strip code was entered into the monitor, and
confirm that the strips have been stored properly. Repeat the test if you believe the strips are
okay. If you continue to get a “LO” or “HI” QC message, make note of the information on the
monitor display and call Technical Support for assistance.
Monitor Display
Mode
INR + QC
or
INR + QC + PT
Example Low
QC1 Display
Example High
QC1 Display
Monitor Display
Mode
INR + QC
or
INR + QC + PT
Example Low
QC2 Display
Example High
QC2 Display
Monitor Display
Mode
INR
or
INR + PT
Example Low
QC2 Display
Example High
QC2 Display