Invacare® TDX® SP2 Series
DANGER!
Risk of Injury, Damage or Death
Overinflation of tires may cause tires to explode.
– Inflate tire to the proper tire pressure (P.S.I. /
kilopascals) listed on the side wall of the tire.
– Only use wheelchair with tires at proper
pressure.
– The wheels and tires should be checked
periodically for cracks and wear and should be
replaced if necessary.
9.2.1 Inspect/Adjust Weekly
q
Ensure that the casters are free of debris.
q
Inspect tires for flat spots and wear.
q
Inspect all fasteners.
q
Inspect TRRO fasteners and hardware.
q
Ensure proper operation of powered functions
(Example: drive, seating and legrests).
9.2.2 Inspect/Adjust Monthly
q
Clean upholstery and armrests.
q
Clean dirt and lint from axles.
q
Clean dirt and lint from bearings.
q
Ensure that the casters are free of debris.
q
Inspect seat positioning strap for any signs of wear.
Ensure buckle latches. Verify hardware that attaches
strap to frame is secure and undamaged. Replace if
necessary.
9.2.3 Inspect/Adjust Periodically
•
Ensure wheelchair rolls straight (no excessive drag or
pull to one side).
•
Inspect all operator (user/attendant) adjustable
fasteners including the back pan, back cane and angle
adjustment fasteners, and the arm support, flip back
and height adjustment fasteners. Ensure fasteners are
securely tightened.
•
Inspect TRRO fasteners and hardware.
•
Ensure clothing guards are secure.
•
Ensure arms are secure but easy to release and
adjustment levers engage properly.
•
Ensure adjustable height arms operate and lock securely.
•
Ensure upholstery does not have any rips or tears.
•
Armrest pad sits flush against arm tube.
•
Ensure that the casters are free of debris.
•
Inspect foam handgrips for damage. If damaged, have
them replaced by a qualified technician.
•
Check center mount front riggings for loose fasteners.
Replace /tighten if necessary.
•
Check that all labels are present and legible. Replace
if necessary.
9.3 Service Inspection
Every six months take your wheelchair to a qualified
technician for a thorough inspection and servicing.
Service inspections MUST be performed by a
qualified technician.
66
60101831-A