IMPORTANT!
Wrong fluids or methods can harm or damage
the product.
– All cleaning agents and disinfectants used must
be effective, compatible with one another and
must protect the materials they are used to
clean.
– Never use corrosive fluids (alkalines, acid etc.)
or abrasive cleaning agents. We recommend
an ordinary household cleaning agent such as
dishwashing liquid, if not specified otherwise in
the cleaning instructions.
– Never use a solvent (cellulose thinner, acetone
etc.) that changes the structure of the plastic
or dissolves the attached labels.
– Always make sure that the product is
completely dried before taking into use again.
For cleaning and disinfection in clinical or long-term
care environments, follow your in-house procedures.
9.6.2 Cleaning Intervals
IMPORTANT!
Regular cleaning and disinfection enhances
smooth operation, increases the service life and
prevents contamination.
Clean and disinfect the product
– regularly while it is in use,
– before and after any service procedure,
– when it has been in contact with any body
fluids,
– before using it for a new user.
9.6.3 Cleaning
IMPORTANT!
– The product does not tolerate cleaning in
automatic washing plants, with high-pressure
cleaning equipment or steam.
IMPORTANT!
Dirt, sand and seawater can damage the bearings
and steel parts can rust if the surface is damaged.
– Only expose the wheelchair to sand and
seawater for short periods and clean it after
every trip to the beach.
– If the wheelchair is dirty, wipe off the dirt as
soon as possible with a damp cloth and dry
it carefully.