8 Troubleshooting
8.1 Troubleshooting Table
SYMPTOMS
PROBLEM
SOLUTION
Emergency stop activated
Check that the emergency stop is not
activated. Refer to
emergency stop or lowering feature, page
18
.
System does not have power or battery
is discharged
Charge the battery or use emergency
battery. Refer to
.
Ceiling hoist does not respond to the hand
control’s buttons and up/down buttons
on the ceiling hoist are not responding.
Fuse is blown
The fuse needs to be replaced. Contact
an Invacare representative or specialized
dealer for service.
Hand control not connected properly
Check the hand control plug. Refer to
Inserting hand control plug, page 13
Ceiling hoist does not respond to the hand
control’s buttons, but up/down buttons
on the ceiling hoist are responding.
Hand control defective
Replace the hand control.
Battery nearly discharged
Charge the battery or use emergency
battery. Refer to
.
Lifting straps will only move down and
not up.
Load is too high
Reduce the load. The hoist is equipped
with an over-current safety system.
Hand control defective
Replace the hand control.
Lifting straps will only move up and not
down.
Loose strap function activated
Contact an Invacare representative or
specialized dealer for service.
Ceiling hoist speed is the same with and
without load.
Non-load quick movement function
activated permanently
Contact an Invacare representative or
specialized dealer for service.