1.
Insert the plug of the external back-up battery
A
into
aperture
B
underside of the hoist. The external back-up
battery should hang freely during patient transfer.
2.
Operate the ceiling hoist as described in chapter
Operating the ceiling hoist, page 14
.
3.
Recharge the external back-up battery in order to have
it fully charged for the next usage.
Charging the external battery:
1.
Place the plug in the charger socket (a special adapter
is necessary).
2.
The LED light turns green to indicate the unit is fully
charged (approx. charging time 2 hours).
5.6 Emergency stop
The emergency stop function is used to stop the patient
from lowering or lifting if the ceiling hoist does not stop
or react to the hand control.
1.
Pull and release the red emergency strap
A
underside
of the hoist
B
to activate the emergency stop.
The ceiling hoist stops any lifting or lowering movement
immediately.
IMPORTANT!
When the red emergency strap has been pulled,
the hoist functions, including the end stop, are
deactivated. The blue indicator light turns off.
– To activate the hoist functions again, reset
the emergency stop. See
emergency stop or lowering feature, page 18
1586745-C
17