Troubleshooting
Flash code
Fault
Consequence for the
scooter
Comments
5
Brake failure
Stops driving
•
Ensure that the disengaging lever is in the engaged
position.
•
There is a defect in the braking coil or in the cabling.
Check the magnetic brake and cabling for open
or short-circuited circuitry. Contact your Invacare
provider.
No neutral position when
switching scooter on.
Stops driving
•
Drive lever is not in neutral when the keyswitch was
turned. Put the drive lever in neutral, turn the power
off and then turn on again.
•
It may be necessary to replace the drive lever. Contact
your Invacare provider.
6
Circuit board failure
Stops driving
•
Circuit board got damaged while connecting or
disconnecting the charger to the charger socket.
•
Test resistance of drive lever to eliminate a defective
lever.
•
It may be necessary to replace the circuit board or
drive lever. Contact your Invacare provider.
7
Fault in speed
potentiometer
Stops driving
•
The drive lever controls could be faulty or
incorrectly connected. Check the cabling for open or
short-circuited circuitry.
•
Potentiometer is not correctly adjusted and must be
replaced. Contact your Invacare provider.
8
Motor voltage error
Stops driving
•
The motor or its cabling is defective. Check the cabling
for open or short-circuited circuitry.
1603456-D
59
Содержание orion pro