MicroVista Camera User Manual
Imaging Support: [email protected] http://www.mosir950.com 408-496-2019
7
To calculate the readout time of a particular window, use the following equation:
The following table lists readout times and frame rates for some common window
sizes:
Window Size
Readout Time
Frame Rate
1280 x 1024
33327 µs
30 Hz
640 x 512
9216 µs
109 Hz
320 x 256
2746 µs
364 Hz
160 x 128
908 µs
1102 Hz
80 x 64
337 µs
2963 Hz
t
read
= ((col
end
- col
start
+ 1)
×
t
pixel
+ t
reset
)
×
(row
end
- row
start
+1)
t
reset
= 152
×
t
pixel
= 3.4545 µs
t
pixel
= ----------- = 0.0227 µs
1
44 MHz