120-00011 Rev. New
Halimeter
PLUS User Manual
7/15/2020
12
indicator as shown on the right. The instrument can be charged with power on or off.
For fastest charging of the Halimeter
PLUS do not connect any other USB devices
to the charger.
NOTE:
The internal batteries must have a charge to maintain the sensor bias voltage
and minimize sensor warmup time. It is recommended you keep the instrument
connected to the power supply for long term storage or extended use.
The Setup Menu shown on the right offers access
to parameters that may need to be changed for the
user’s particular application as well as maintenance
functions. This menu is detailed below.
Flow rate
– cc / min
– Nominal setting =
200 cc / min
.
Do not change this setting.
Sampling Mode
– Nominal setting =
HAL3
. This setting controls the SAMPLE
MODE procedure with 4 options as follows:
•
Continuous
- Sample is drawn continuously.
•
HAL1
-
Sample is drawn in a single cycle of 3-minute hold followed by a
35 second breath sample.
•
HAL2
- Sample is drawn in 2 sample cycles as noted in HAL1.
•
HAL3
- Sample is drawn in 3 sample cycles as noted in HAL1. This is the
default Halimeter
PLUS sampling mode.
Datetime
– Set the date and time according to the user’s time zone.
Breath Calibration
– Provides for user calibration using a NORMAL BREATH
sample. See section 7 for details on calibration.
ECS Calibration
–This feature is intended for non-Halimeter applications and will
not be used on the Halimeter
PLUS.
Factory Setup
– A password protected setup sub-menu for factory setup only.
3.6 SETUP MENU