Display Filter Command (DF)
The Display Filter (DF) command sets the filter type and level. The format is
DF(type)(level), where type is either 1 for type I filters, or 2 for type II filters, level must
be 1, 2, 3, or 4 depending on the level of filtering desired. (4 is most filtering and
slowest rise time).
Example:
@123DF13
Acknowledge:
@ 123 Filter is Type I Level 3
Display Decimal Command (DD)
The Display Decimal (DD) command sets the maximum number of decimal digits that
are shown for Load, Peak, and Valley. The format is DD(channel)(number), where
channel is A or B and the number must be between zero and 5.
Example:
@123DDA0
Acknowledge:
@ 123 Channel A shows 0 decimal digits
Display Position
OPT
Decimal Command (DP)
The Display Position Decimal (DP) command sets the maximum number of decimal
digits that are shown for the position. The format is DP(number). The number must
be between zero and 5.
Example:
@123DP4
Acknowledge:
@ 123 position shows 4 decimal digits
Display Count By Command (DC)
The Display Count by (DC) command sets the
smallest number that the display will count by
for the Load or Torq, Peak, and Valley
measurements. The format is
DC(channel)(code) where the code is selected
from the table at the right.
Example:
@123DCA3
Acknowledge:
@123 Channel A counts by 10
Display Count by Command
Code
Count By
0
1
1
2
2
5
3
10
4
20