Check battery and connect to charging.
Check the load weight is seated properly.
Check the load cell connections and connecters.
Incorrect value:
Calibration error.
Calibrate again with exact calibration mass weights.
Check the weight sample is lying proper and avoids touching the cover or
surface.
Check power supply and battery.
Check load cell connections and connectors.
Can not use scale full capacity:
Before weighing make sure zero indication is showing and scale is empty.
Check the weighing mode.
Check the load cell is fitted proper and touching housing or hitting
somewhere.
Calibrate again with exact calibration mass weights.
Load cell damaged. Check load cell connections and connectors.
Main PCB problem.
Battery not charging:
Mains voltage problem. Check the power supply voltage and adaptor voltage.
Charging circuit failure.
Battery failure, check the battery connections