- 25 -
10. TROUBLE SHOOTING
No Display:
Mains power is turned off or power supply not plugged proper.
Power supply faulty.
Internal Battery is not charged.
Check On/OFF switch is turned on and faulty or not.
Check the PCB power connecters and cable.
Display is Blank after the self test / Err stuck:
Unstable weight.
Check the platform is installed correctly.
Try again to turning on.
Check the load cell is not touching any where.
Load cell is damaged. Check the load cell connections and all.
OL or( -------) appear the display:
Maximum capacity exceeded.
Power supply faulty. Check all power cables and connecters.
Calibrate again with correct calibration weights.
Load cell damaged. Check load cell connections.
(------) or Lo:
Weight is below permissible limit.
Check the pan installed correctly.
Calibrate again with correct calibration weights.
Power Supply faulty. Check all power cables and connecters.
Load cell damaged. Check load cell connections and connecters.
Try to turn on again.
Unstable display:
Check the pan is seated proper and touching some where
Check any vibrations, noises, sudden temperature changes.
Check power supply.
Содержание QHD+15
Страница 20: ... 20 Send PLU text data to be stored with current TARE and U W values ie SPLU01 Parts CR ...
Страница 31: ... 31 12 CIRCUIT DIAGRAM ...
Страница 32: ... 32 ...
Страница 33: ... 33 13 DRAWING ...