8. Inspect the equipment to make sure there is no visible organic
debris or residue from cleaning agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness or excess liquid
is detected, dry with a soft clean towel.
• Medically compressed air can be used if needed.
Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions cause
damage to the devices.
• Channels and crevices found on these devices require particular
attention during cleaning.
• Pay special attention to the water quality used throughout
reprocessing. Hard water can damage the surface of the
equipment. Avoid using hard water. Instead use purified water
unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on these devices and
is not recommended for routine use.
• It is important to have Integra NeuroSpecialists perform routine
inspections (twice yearly is recommended).
• See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH Detergent
Method
1. Prepare equipment for cleaning by disassembling removable parts
and loosening connections.
2. Rinse equipment in warm water before placing into washer.
3. Load device into the washer and place small parts in container
or tray inside the washer unit in order to avoid losing small
components.
NOTE
– Load devices carefully into washer in order to
avoid collision.
4. Follow the instructions listed below and set washer machine to
these exact parameters:
Phase
Time
(Min.)
Water
Temperature
Detergent and
Concentration
Pre-wash 1 4:00
Cold water
N/A
Enzyme
Wash
4:00
Hot water
Neutral pH enzymatic (e.g.
Endozime® AW Triple Plus
with APA, Ratio 1:128)
Wash 1
10:00
60.0ºC
(140º F)
Neutral pH detergent
(e.g. Renu-Klenz™ , Steris,
Ratio 1:256)
Rinse 1
0:30
Hot water
N/A
Thermal
Rinse**
2:00
82.2ºC
(180º F)
N/A
** Optional phase for disinfection of components – minimum water
temperature as indicated or per manufacturer specifications for the
thermal rinse cycle.
NOTE
– Any deviation from this guideline could result in damage to
the equipment as well as improper cleaning results.
Rinse with purified water. Do not perform if parameters cannot
be achieved.
5. Remove from washer and dry completely if needed.
6. Inspect equipment to make sure there is no visible organic debris
or residue from cleaning agent.
Repeat process if any visible soil is detected.
The life expectancy of the MAYFIELD products is expected to be as
long as 7 years.
Maintenance and Care
To ensure proper function and to extend the life and performance of
the equipment, Integra LifeSciences recommends the following:
Recommended Action
Recommended Frequency
Return the device to the Integra
LifeSciences Repairs department for
detailed inspection and servicing.
Once / year
Request that Integra NeuroSpecialists
perform routine inspections of the
device.
Twice / year
In the absence of proper care and servicing of the device, negative
effects may be seen after repeated processing over time which may
lead to reduced performance.
Contact information: See the Service and Repair section for contact
information on how to return your device for periodic servicing and
to request periodic inspections.
See Inspection and /or Service notes section for routine checks to be
performed on the device.
NOTE
– Any serious incident that has occurred in relation to the
device for the user and/or the patient should be reported to the
manufacturer and the competent authority of the member state in
which the user and/or patient is established.
Device Disposal
NOTE
– Follow hospital procedures for disposal of this device.
Содержание MAYFIELD A2101
Страница 2: ...THIS PAGE IS INTENTIONALLY LEFT BLANK...
Страница 10: ...FR Fran ais 8 CETTE PAGE EST LAISS E INTENTIONNELLEMENT VIDE...
Страница 11: ...FR Fran ais 9 Unit de base Ultra MAYFIELD A2101 et A2102 Mode d emploi...
Страница 18: ...IT Italiano 16 QUESTA PAGINA STATA LASCIATA INTENZIONALMENTE IN BIANCO...
Страница 19: ...IT Italiano 17 MAYFIELD Unit Ultra Base A2101 A2102 Manuale di istruzioni...
Страница 26: ...DE Deutsch 24 DIESE SEITE WURDE ABSICHTLICH FREI GELASSEN...
Страница 27: ...DE Deutsch 25 MAYFIELD Ultra Basiseinheit A2101 und A2102 Bedienungsanleitung...
Страница 34: ...ES Espa ol 32 P GINA DEJADA INTENCIONALMENTE EN BLANCO...
Страница 35: ...ES Espa ol 33 Unidad Ultrabase MAYFIELD A2101 y A2102 Manual de instrucciones...
Страница 42: ...NL Nederlands 40 DEZE PAGINA IS BEWUST LEEG GELATEN...
Страница 43: ...NL Nederlands 41 MAYFIELD Ultra basisunit A2101 A2102 Instructiehandleiding...
Страница 50: ...SV Svenska 48 DENNA SIDA HAR AVSIKTLIGEN L MNATS BLANK...
Страница 51: ...SV Svenska 49 MAYFIELD Ultra basenhet A2101 A2102 Instruktionshandbok...
Страница 58: ...SV Svenska 56 DENNA SIDA HAR AVSIKTLIGEN L MNATS BLANK...
Страница 59: ...SV Svenska 57 DENNA SIDA HAR AVSIKTLIGEN L MNATS BLANK...