Key in calibration. 



Once the 

KEY IN 

option has been selected from the Calibration menu, 

4 MA 

will appear on the display and toggle  

with the default display value for 4mA input (default = 0). Use the 

U

 and 

D

 buttons to adjust the 4mA display value 

as desired, and then press 

S

 to continue. 



20 MA

 will now appear on the display and toggle with the default display value for 20mA input (default = 

10,000).  
Use the 

U

 and 

D

 buttons to adjust the 20mA display value as desired. Then press 

S

 to save the key in calibration  

values and return to the operational display. 

 

2 point calibration. 



Once the 

2POINT

 option has been selected from the Calibration menu, 

Point1

 will appear on the display and toggle 

with the default display value for the first calibration point (default = 0). Apply the Point 1 input signal to the LPI-LCD-6 
display. Use the 

U

 and 

D

 buttons to adjust the display value as required, and then press 

S

 to continue. 



Point2

 will now appear on the display and toggle with the default display value for the second calibration 

point  
(default = 10,000). Apply the Point 2 input signal to the LPI-LCD-6 display. Use the 

U

 and 

D

 buttons to adjust the  

display value as required, and then press 

S

 to save the 2 point calibration values and return to the operational  

display. 



If calibration was successful, the display will return to operating mode. 



If 

ERROR

 appears on the display, then the calibration has not been saved. The most common cause of a  

calibration error is an insufficient change in input signal between the 2 points. Press 

S

 to exit to the operational 

display, or the 

B

 button to go back to the calibration menu. See Troubleshooting (below) for more information 

on this error. 

 

Reset. 



If you select 

RESET

, there is nothing further to configure. The LPI-LCD-6 display will restore the factory calibration 

and return to operational display.  

 

 

Backlight. 



Enter the 

B.LIGHT 

menu (see page 6 under ‘Main Menu’). 



The currently selected backlight time will appear. This is the amount of time that the backlight will remain on for after 
the 

L

 button is pressed (default = 3 seconds). Use the 

U

 and 

D

 buttons to adjust the time from 1~9 seconds, and 

then press 

S

 to save and return to the operational display. 

 

Note

:  The backlight charges off the input signal and is not intended to be used for extended periods of time. It may  

appear dimmer if used repeatedly, or if used immediately after the LPI-LCD-6 display has been connected. 

 

 

Contrast. 



Enter the 

CNTRST

 menu (see page 6 under ‘Main Menu’). 



The currently selected screen contrast will appear on the display. Use the 

U

 and 

D

 buttons to adjust the contrast  

between 

CNT. 1

 and 

CNT. 6

, and then press 

S

 to save and return to the operational display. 

 

 

Troubleshooting. 



ERROR

 - This error appears during 

2POINT 

calibration and occurs when there is insufficient change in input signal 

between the 2 points or when the range specified by the controller exceeds the LPI-LCD-6’s internal limitations. To 

correct this error, please check your input signal and connections, and try calibrating again. Your display range may 

have a maximum gain of 100 (i.e. a 2mA change on the input must not exceed 200,000 change in display counts). 



ER.FLSH

 - A flash error normally occurs when there is an unexpected loss of power to the instrument during setup.  

If this error occurs, all settings will be restored to their factory defaults. To clear the error message, press the 

S

 

button. Your previous setup cannot be recovered - please configure the instrument again. 



ER. ADC 

- If this error occurs, then there is a problem with the LPI-LCD-6’s  onboard A/D converter. Pressing the 

S  

button on the front panel will clear the error message temporarily, however the display may stop responding to  

changes in input signal. If you get this error, try powering the unit off and then on again. If the fault persists, contact 

Intech Instruments Ltd for support. 



ER.FCAL 

- If a factory calibration error occurs, then the instrument's factory calibration has been lost. Pressing the 

S

 

button on the front panel will clear the error message temporarily. Please contact Intech Instruments Ltd for support. 

Содержание LPI-LCD-6

Страница 1: ...1 LPI LCD 6 4 20mA Panel Display Installation Guide...

Страница 2: ...rminal Layout Page 5 LPI LCD 6 Wiring Diagrams Page 5 LPI LCD 6 Buttons Display Page 5 Display Screen Page 5 LPI LCD 6 Setup Calibration Page 6 Main Menu Page 6 Changing the Decimal Place Page 6 Setti...

Страница 3: ...No liability will be accepted for errors omissions or amendments to this specification Technical data are always specified by their average values and are based on Standard Calibration Units at 25 C u...

Страница 4: ...kers to assist with installation Apply the front sticker if you are drilling from the front of the panel and back sticker if you are drilling form the back of the panel After drilling remove the insta...

Страница 5: ...out saving changes Display Screen The screen can display 1 6 characters to reflect the condition of your 4 20 input in normal operating mode The character on the right may be set up to display units a...

Страница 6: ...n after the six display digits If you selected CUSTOM use the U and D buttons to select any one letter from A Z and then press S to save and return to the operational display the custom unit you choos...

Страница 7: ...ill appear This is the amount of time that the backlight will remain on for after the L button is pressed default 3 seconds Use the U and D buttons to adjust the time from 1 9 seconds and then press S...

Страница 8: ...current signals Austral Standard Cables B5102ES is recommended For 3 wire transmitters Austral Standard Cables B5103ES is recommended 4 It is recommended that you do not ground current loops and only...

Отзывы: