6
Changing the Decimal Place.
Enter the
D.POINT
menu (see above ‘Main Menu’).
Your current decimal point position will appear on the display - either:
X.XXXX
,
X.XXX
,
X.XX
,
X.X
or
X
. Use the
U
and
D
buttons to shift the decimal point left or right, and then press
S
to save and return to the operational display.
Note: Adjusting your decimal point position will not scale your calibration.
4 decimal places (
X.XXXX
) are only available when custom units are not used.
Setting the Rounding.
Enter the
ROUND
menu (see above ‘Main Menu’).
The currently selected display rounding will appear. Use the
U
and
D
buttons to select:
OFF
,
XXXX0
, or
XXX00
,
and then press
S
to save and return to the operational display.
Note: Rounding is quoted in display counts and is not influenced by decimal point position.
Setting the Units.
Enter the
UNITS
menu (see above ‘Main Menu’).
The currently selected display units will appear. Use the
U
and
D
buttons to select:
OFF
,
K.G.
,
LB.
,
TON.
, or
CUSTOM
, and then press
S
to save.
If you selected
OFF
,
K.G.
,
LB.
or
TON
, there is nothing further to configure. After pressing
S
the display
will return to normal operating mode.
OFF
means no units will be displayed. By selecting
K.G.
,
LB.
or
TON.
,
then the respective units will be displayed in the units area at the right edge of the screen after the six display
digits.
If you selected
CUSTOM
, use the
U
and
D
buttons to select any one letter from A~Z, and then press
S
to
save and return to the operational display. the custom unit you choose will appear in the sixth character
position on the display which means your usable display counts will be reduced to five digits.
Calibration.
Enter the
CAL.
menu (see above ‘Main Menu’).
The currently selected calibration method will appear on the display. Use the
U
and
D
buttons to select a
calibration method:
KEY IN
,
2POINT
, or
RESET
, and then press
S
to continue.
KEY IN
scales the display using high and low display values, and relies on the factory set calibration.
2POINT
calibrates the unit using live input signals. A calibrator or other external signal source is
required for this procedure.
RESET
will restore the factory settings.
In the calibration adjustment screens, press and hold the
U
or
D
button to adjust the display in 10s, 100s,
1000s etc. Press the
U
and
D
buttons at the same time to restore to the default value.
Pressing the
B
button at any stage before calibration has completed will exit the calibration menu without
saving any of your calibration values.
LPI-LCD-6 Setup & Calibration.
Main Menu.
From the operational display area, press the
S
‘MENU’ button .
‘D.POINT’ will appear on the display. Scroll through the menu items
using the ‘UP’ and ‘Down’ buttons
U D
. Press
S
to enter the
selected menu, or the ‘BACK’ button
B
to go back to your
operational display.