![Instrutech The Hornet IGM402 Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/instrutech/the-hornet-igm402/the-hornet-igm402_user-manual_2070238053.webp)
Instruction Manual IGM402 Hornet
InstruTech
Page 51
CAUTION!
Do not assume this data applies to other convection gauges, which may or may not be the same.
The table below shows the convection gauge displayed readings at various pressures for several commonly used
gas types:
Displayed pressure readings for convection gauge type devices versus true pressure for selected gases
True
Pressure
(Torr)
N
2
Ar
He
O
2
CO
2
Kr
Freon12 Freon22
D
2
Ne
CH4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
1.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
2.00E-4
5.00E-4
5.00E-4
5.00E-4
5.00E-4
5.00E-4
5.00E-4
3.00E-4
5.00E-4
5.00E-4
5.00E-4
5.00E-4
5.00E-4
1.00E-3
1.00E-3
7.00E-4
8.00E-4
1.00E-3
1.10E-3
4.00E-4
1.50E-3
1.50E-3
1.30E-3
7.00E-4
1.70E-3
2.00E-3
2.00E-3
1.40E-3
1.60E-3
2.00E-3
2.30E-3
1.00E-3
3.10E-3
3.10E-3
2.40E-3
1.50E-3
3.30E-3
5.00E-3
5.00E-3
3.30E-3
4.00E-3
5.00E-3
4.40E-3
2.30E-3
7.60E-3
7.00E-3
6.00E-3
3.50E-3
7.70E-3
1.00E-2
1.00E-2
6.60E-3
8.10E-3
9.70E-3
1.10E-2
4.80E-3
1.47E-2
1.35E-2
1.21E-2
7.10E-3
1.53E-2
2.00E-2
2.00E-2
1.31E-2
1.61E-2
1.98E-2
2.22E-2
9.50E-3
2.99E-2
2.72E-2
2.43E-2
1.41E-2
3.04E-2
5.00E-2
5.00E-2
3.24E-2
4.05E-2
4.92E-2
5.49E-2
2.35E-2
7.25E-2
6.90E-2
6.00E-2
3.48E-2
7.72E-2
1.00E-1
1.00E-1
6.43E-2
8.20E-2
9.72E-2
1.07E-1
4.68E-2
1.43E-1
1.36E-1
1.21E-1
7.00E-2
1.59E-1
2.00E-1
2.00E-1
1.26E-1
1.65E-1
1.94E-1
2.10E-1
9.11E-2
2.75E-1
2.62E-1
2.50E-1
1.41E-1
3.15E-1
5.00E-1
5.00E-1
3.12E-1
4.35E-1
4.86E-1
4.89E-1
2.17E-1
6.11E-1
5.94E-1
6.87E-1
3.59E-1
7.81E-1
1.00E+0
1.00E+0
6.00E-1
9.40E-1
9.70E-1
9.50E-1
4.00E-1
1.05E+0
1.04E+0
1.55E+0
7.45E-1
1.60E+0
2.00E+0
2.00E+0 1.14E+0 2.22E+0 1.94E+0 1.71E+0
7.00E-1
1.62E+0
1.66E+0
4.13E+0 1.59E+0 3.33E+0
5.00E+0
5.00E+0 2.45E+0 1.35E+1 4.98E+0 3.34E+0 1.28E+0
2.45E+0
2.62E+0
2.46E+2 5.24E+0 7.53E+0
1.00E+1
1.00E+1 4.00E+0
OP
1.03E+1 4.97E+0 1.78E+0
2.96E+0
3.39E+0
OP
2.15E+1 2.79E+1
2.00E+1
2.00E+1 5.80E+0
OP
2.23E+1 6.59E+0 2.29E+0
3.32E+0
3.72E+0
OP
5.84E+2 3.55E+2
5.00E+1
5.00E+1 7.85E+0
OP
7.76E+1 8.22E+0 2.57E+0
3.79E+0
4.14E+0
OP
OP
8.42E+2
1.00E+2
1.00E+2 8.83E+0
OP
2.09E+2 9.25E+0 2.74E+0
4.68E+0
4.91E+0
OP
OP
OP
2.00E+2
2.00E+2 9.79E+0
OP
2.95E+2 1.23E+1 3.32E+0
5.99E+0
6.42E+0
OP
OP
OP
3.00E+2
3.00E+2 1.13E+1
OP
3.80E+2 1.69E+1 3.59E+0
6.89E+0
7.52E+0
OP
OP
OP
4.00E+2
4.00E+2 1.35E+1
OP
4.85E+2 2.24E+1 3.94E+0
7.63E+0
8.42E+0
OP
OP
OP
5.00E+2
5.00E+2 1.61E+1
OP
6.04E+2 2.87E+1 4.21E+0
8.28E+0
9.21E+0
OP
OP
OP
6.00E+2
6.00E+2 1.88E+1
OP
7.30E+2 3.64E+1 4.44E+0
8.86E+0
9.95E+0
OP
OP
OP
7.00E+2
7.00E+2 2.18E+1
OP
8.59E+2 4.61E+1 4.65E+0
9.42E+0
1.07E+1
OP
OP
OP
7.60E+2
7.60E+2 2.37E+1
OP
9.41E+2 5.39E+1 4.75E+0
9.76E+0
1.11E+1
OP
OP
OP
8.00E+2
8.00E+2 2.51E+1
OP
9.97E+2 5.94E+1 4.84E+0
9.95E+0
1.14E+1
OP
OP
OP
9.00E+2
9.00E+2 2.85E+1
OP
OP
7.95E+1 4.99E+0
1.05E+1
1.20E+1
OP
OP
OP
1.00E+3
1.00E+3 3.25E+1
OP
OP
1.11E+2 5.08E+0
1.11E+1
1.27E+1
OP
OP
OP
Values listed under each gas type are in Torr units.
OP = Overpressure
Example: If the gas is argon (Ar), when the true pressure is 100 Torr, the gauge will read about 9 Torr.