Instruction Manual IGM402 Hornet
InstruTech
Page 44
6.3
Convection gauge analog output - LOG-LINEAR
ANALOG
TYPE in
SETUP CG1 or SETUP CG2 sub
menu of
SETUP CGS
menu is set to
LOG-LINEAR.
1 to 8 Vdc, 1 V/decade (N
2
/air only)
This setting provides a log-linear analog output for convection gauges 1 and 2. The analog output is
proportional to pressure measurements from either of the convection gauges 1 or 2 as described below:
A) The log-linear output signal and pressure are related by the following
formulas when units of measurement is in
Torr
and
mbar:
P = 10
(volts – 5)
V = log
10
(P) + 5
Where P is the pressure in Torr or mbar, and V is the output signal in volts.
B) The log-linear output signal and pressure are related by the following
formulas when units of measurement is in
pascals:
P = 10
(volts – 3)
V = log
10
(P) + 3
Where P is the pressure in pascals, and V is the output signal in volts.
The following chart shows the graphical results of table and formulas
above for pressure measurements in Torr. Pressure is plotted on the X- axis
with a log scale; the output signal is plotted on the Y-axis on a linear scale.
Log-Linear Analog Output
(N
2
/air only)
Pressure
(Torr)
Voltage
1.0E-04
1.000
2.0E-04
1.301
5.0E-04
1.699
1.0E-03
2.000
2.0E-03
2.301
5.0E-03
2.699
1.0E-02
3.000
2.0E-02
3.301
5.0E-02
3.699
1.0E-01
4.000
2.0E-01
4.301
5.0E-01
4.699
1.0E+00
5.000
2.0E+00
5.301
5.0E+00
5.699
1.0E+01
6.000
2.0E+01
6.301
5.0E+01
6.699
1.0E+02
7.000
2.0E+02
7.301
3.0E+02
7.477
4.0E+02
7.602
5.0E+02
7.699
6.0E+02
7.778
7.0E+02
7.845
7.6E+02
7.881
8.0E+02
7.903
9.0E+02
7.954
1.0E+03
8.000
Log-Linear Analog Voltage to Pressure
0.000
1.000
2.000
3.000
4.000
5.000
6.000
7.000
8.000
9.000
1.0E-04
1.0E-03
1.0E-02
1.0E-01
1.0E+00
1.0E+01
1.0E+02
1.0E+03
Pressure (Torr)
O
utp
ut
S
ign
al
(
V
ol
ts
dc
)