Chapter 4: Parts Replacement
54
M47-17036-EN
Replacement of fuses
The system is equipped with fuses as outlined in
These fuses should only need replaced when they are blown. Problems that indicate a blown fuse are
described in
. If you believe a fuse could be blown, perform
the following check to pinpoint which fuse is blown:
1. Start up the system according to the system startup procedure provided in the System Operating
Instructions (supplied separately).
2. Check the indicators on the
SERVICE
display of the 59 Series control unit (see page
should be lit. If the indicators are not lit, then the power entry module fuses could be blown. Visually
inspect the fuse. Follow the procedure outlined under
“Inspect/replace 59 Series power entry
to access and locate the fuse. If the fuse is blown, replace it.
Information for the fuse can be found in
Table 10. Fuse information.
Fuse Description
Specifications
Instron
Part Number
Quantity
Fuses in power entry module of 59 Series control
unit
5 amp,
1/4D x 1-1/4LG, 250Vac
(equivalent metric size is
6.3D x 32LG)
300-8905-9159
2