![Instron 1000KPX-J3 Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/instron/1000kpx-j3/1000kpx-j3_system-concepts-manual_2070124047.webp)
47
Frame
Product Support:
Troubleshooting
In the event that problems arise during operation of the system, refer to
for help in determining
the specific problem and its solution. If the problem cannot be determined through the chart, contact
your local Instron Services department as directed on page
. Another option would be to check the
listing of Frequently Asked Questions (FAQs), available on the Instron website (
www.instron.com
), for a
description/solution to your problem.
Table 9. Troubleshooting the KPX system.
No.
Possible Indications
Possible Problem
Solution
1
•
The indicators on the
SERVICE
display do not light when the system
disconnect switch is turned on
•
Frame power is lost for no apparent
reason
•
Test aborts and HPS shuts down
•
Frame will not enable
•
An interlock error message appears
in the controlling software
Frame and/or 59 Series control
unit are not receiving power
•
Check incoming power supply for
both frame and control unit
A fuse could be blown
•
Check all system fuses, refer to
2
•
HPS will not start
Always verify that:
•
Computer is on
•
59 Series control unit is on
•
Controlling software is running
•
Frame is enabled
•
A test procedure/method is open
•
The load transducer is calibrated
Emergency Stop
engaged
Disengage
Emergency Stop
button,
enable frame and start HPS
System disconnect switch not ON
( | )
Turn the disconnect switch to ON ( | ).
Overload protect circuit may have
tripped
For assistance, contact your local Instron
Services department as directed on
page
If none of the above appear to be the problem:
•
For Partner systems, check the Status display for any abnormal
conditions or check the On-line Help for other helpful information
•
For Bluehill systems, check the Status Log for any abnormal conditions.
Click the Load Frame icon; the Status Log appears on the bottom portion
of the Load Frame tab.
Содержание 1000KPX-J3
Страница 16: ...Chapter 1 Introduction 16 M47 17036 EN This page is intentionally blank...
Страница 52: ...Chapter 3 Additional System Details 52 M47 17036 EN This page is intentionally blank...
Страница 56: ...Chapter 4 Parts Replacement 56 M47 17036 EN This page is intentionally blank...
Страница 57: ......
Страница 58: ...Product Support www instron com...