Specifications
67
200J of energy at 175R impedance.
Alarm system
Eazy PRO’s alarm system is made of a single alarm, as shown in the table below:
ALARM CONDITION
The device is ready to apply the shock treatment to the patient using the
energy previously established by either the algorithm (AED mode) or the
operator (manual override mode), waiting only for the operator to press the
button.
ALARM PRIORITY
High.
VISUAL ALARM SIGNAL
Button blinks in red, at 2Hz, and at a 50% duty cycle.
VERBAL ALARM SIGNAL
Voice indication with the following text: “push the flashing button to treat the
patient ”.
SOUND LEVEL
Configurable between 65 and 85 dB(A), through PC software application:
SOFTDEA.
ALARM SIGNAL GENERATION
DELAY
Lower than 5s.
OPERATOR’S POSITION
The operator must be facing the front side of the device, at a maximum
distance of 50cm.
Pausing and turning off the alarm signals:
Following the standard NBR IEC 60601-2-4, it is not possible to turn-off or pause
the alarm signals for the device’s ALARM CONDITION.