The Equipment
22
ATTENTION: remember to check the status of the operational status indi-
cator at least every 30 days.
SOUND INDICATOR
Along with visual indication, the Eazy PRO emits electronic beeps which can be
identified by the user from a distance and can also be interpreted according to the
frequency, as in the table below:
Number of beeps
Failure description
Two
CPU failure
Three
Failure in defibrillation module
Four
Discharged battery
ATTENTION: the device will not switch on if the battery is discharged
or showing general failure. In this case only the sound warning of the
respective failure will be emitted.