![Instramed EazyPro Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/instramed/eazypro/eazypro_user-manual_2070101021.webp)
User manual |
The Equipment
21
2 - Operational status indicator
The Eazy PRO performs a full auto-test weekly, allowing the user to know the
operational status of the device. This status is informed by a visual indicator (see
picture below), voice messages and sound signals.
The automatic test is also performed when the device is switched on. Should any
problem be found, the voice message “problem with the automatic test” will be
emitted with the visual indicator of failure.
VISUAL INDICATOR
Shows that the device
DOES NOT
HAVE ENOUGH BATTERY CHARGE
TO OPERATE
or has another internal
defect. Charge battery immediately. If
the indicator remains red, please call
technical assistance.
Shows that the device is operating and
ready for use.
• Mains supply LED on means device is connected to the electric current.
• Mains supply LED blinking when in normal use means the battery needs
charging.
In this state the Eazy PRO guarantees at least 3 full energy
discharges.
• Battery LED on means internal battery is being charged. A blinking LED beside
this symbol means full charge.
NOTE: EVEN AFTER THE BATTERY HAS BEEN COMPLETELY CHARGED,
the operational status indicator may continue to show an for some
time.
The display will only change from to when the Eazy PRO performs
the auto-test routine or if the device is turned on/of by the user.