Operating in manual mode
36
Step 2
Remove pads from their wrapping and peel off the film protecting the
adhesive.
Place pads on the patient according to the picture above, keeping adhesive
area in contact with the skin.
This position allows the electric current to circulate from one pad to the other, thus
reaching the whole thoracic cage.
ATTENTION: the area in contact with the pads must be dry.
The presence of too much hair in the contact area may affect scanning. In
this case, shave hair.
ATTENTION: the pads must be applied directly over the skin. DO NOT
place pads over clothes.
ATTENTION: The pads are disposable and for single-use, cannot be reused
under any circumstances.
ATTENTION: After opening the wrapping, the pads should be used within
24 hours.
ATTENTION: In case of use for long periods, the pads should be replaced
every 24 hours.