IP Connect

The IP Connect pack is used in conjunction with the discharging 
hospital. This pack is charged in the same way and has the 
same battery life as the External Battery Pack (see page 8 for 
information). Full training will be given on its use.

Shown across is the icon onscreen confirming connection.

Warnings

Pump will continue to function. Alarm will sound intermittently. 

  Press soft key below the bell sign to silence the alarm

Near End of Infusion

Pump will alarm when nearing the end of the VTBI. 
Time is preset at 20 minutes. This can be changed in 
the configuration menu to volume remaining or time 
remaining, or turned off.

Low External Battery Pack

Alarm will sound 40 minutes before the battery 
is completely depleted. The Internal battery will 
AUTOMATICALLY take over the running of the pump.

  Press the bell sign twice to silence the alarm 

  Change to another External Battery Pack or connect to the mains supply

Low Internal Battery

Alarm will sound 40 minutes before internal battery is depleted. 

  Press bell sign twice to silence the alarm 

  Connect to External Battery Pack or mains supply and change internal  

battery when available

  Ensure internal battery indicator is showing 2 or more blocks

Mini Rythmic

 PN+ 

9

Battery status indicator
Connection status indicator

Connector for the charger

On/off button

Cable connector for the pump

Содержание Mini Rythmic PN+

Страница 1: ...Training videos are also available please contact your hospital for further information Mini Rythmic PN Set up Guide...

Страница 2: ...Hospital Nutrition Team Telephone No Homecare Company Telephone No Current Pump Serial No Inspiration Healthcare 2...

Страница 3: ...e been installed Auto Ramp 25 Steps Continuous Volume Over Time Intermittent Mode Enter Code 0112 1 Purge button used to prime non gravity giving set 2 On off button one button with dual action 3 Soft...

Страница 4: ...confirm all details entered press enter button 4 Pump on Hold if details are correct double press start stop button 5 to start the pump New Patient Programme Auto Ramp Set bag volume press enter butto...

Страница 5: ...press enter button 4 Delay time insert time if required or press enter button 4 Screen will now confirm all details entered press enter button 4 Pump on Hold if all details correct double press start...

Страница 6: ...iming 1 Spike the bag keep filter vertical any air will be eliminated via the filter 2 Press priming button do not use a sharp object until filter is saturated 3 Turn the priming button upside down co...

Страница 7: ...k Mini Rythmic PN Pump Priming Close clamp prior to spiking the bag Load giving set into pump and follow pump programming until pump on hold Hold the pump in a vertical position with the blue clamp at...

Страница 8: ...tery Pack The External Battery Pack should be connected to the pump every time it is used even when priming the giving set The life of a battery pack fully charged when used with a Mini Rythmic PN is...

Страница 9: ...n menu to volume remaining or time remaining or turned off Low External Battery Pack Alarm will sound 40 minutes before the battery is completely depleted The Internal battery will AUTOMATICALLY take...

Страница 10: ...ed Downstream Occlusion Indicates there is a problem between the pump and the patient Press enter button 4 to silence the alarm Rectify the problem and pump will automatically restart Upstream Occlusi...

Страница 11: ...This indicates either the internal door or external casing is not closed properly This alarm will also appear if an attempt has been made to start the pump with no giving set in place Press enter but...

Страница 12: ...in this leaflet are correct at time of going to press The company reserves the right to improve the products shown 500 402 7 2017 05 T 44 0 1455 840 555 E info inspiration healthcare com W inspiratio...

Отзывы: