Mini Rythmic
™
PN+
Depleted External Battery
Press key below bell sign twice to silence the alarm
Change to fully charged External Battery pack
or connect to mains power supply
Air in line
Press button (4) to silence alarm
Message will show ‘Air in line’ setting
Turn pump off; follow hospital protocol
End of Infusion
VTBI has been reached and/or Keep Vein Open is complete.
Press start/stop button (5) to turn pump off
Door open
This indicates either the internal door or external casing
is not closed properly. This alarm will also appear if an
attempt has been made to start the pump with no giving
set in place.
Press enter button (4) to silence the alarm
Ensure doors are closed and giving set is loaded
Press start/stop button (5) twice to start the pump