3
2
Back panel:
1
2
3
4
5
6
7
8
9
10
1. ON/OFF Switch
2. Spare Fuse (250VAC, 800mA)
3. Power Inlet Point
4. Connector for Edge Detector (Optional) 5. Rotary Encoder Input
6. Y
Linear Encoder Input
7. X Linear Encoder Input
8. Printer Interface
9.
Foot Switch connector
10. USB Connector
3
Layout of the Screen:
1
2
3
4
5
6
7
8
10
9
1. Unit
2
.
Measurement Mode
3
.
Coordinate System
4
.
Permenent Storage
5
.
Message Window
6
.
X Axis
7
.
Y Axis
8
.
Angular display
9
.
Scroll Soft Key status
10
.
Soft Key
The setup can be accessed by pressing the followed by key .
The user will be prompted for password. The password is “98563”.
Use [ ]or [ ] key to scroll within the menu.
Setup
24
Point Tolerance
:
Highlight the desired point feature from the result list. Press
{ } key and then Press { } key.
1
Enter the nominal values for X and Y co-ordinates. Also enter the
tolerance zone.
Tolerance zone is the radius within which the point should be
located. If the point is located outside the zone, tolerance fails.
Press { } key then enter
the
nominal
values
for
X
and
Y
co
-
ordinates
.
Also
enter
the
+ /-
values
for
each
nominal
value
.
Limits allow the user to set upper and lower limits on a value.
For example, a nominal value of 10.3250 might have an upper
limit of 10.3260 and a lower limit of 10.3240. For example, if a
part drawing gives a +/- tolerance of 0.005, simply enter 0.005 in
the plus and minus fields.
After set press { } key. The result of tolerance Pass or Fail is
shown for the selected feature.
Use the same way to set Tolerance for line, angle, circle and
distance.
2
Maintenance
1
The projector is a kind of precision optical instrument. Attention
should be paid to its maintenance so as to guarantee the precision
and lengthen life of the instrument. The instrument should be
installed in a clean room, the temperature of which should be
maintained at 20
° ±
5
℃
. The relative humidity of the room should
not exceed 60
﹪
so as to prevent the optical parts and reflecting
mirror from molding and to keep the high quality of the image.