insize ISP-A5000 Скачать руководство пользователя страница 4

3

2

Back panel:

1
2
3

4

5

6

7

8

9

10

1. ON/OFF Switch

 

2. Spare Fuse (250VAC, 800mA)

 

3. Power Inlet Point

 

4.  Connector  for  Edge  Detector  (Optional)  5.  Rotary  Encoder  Input

 

6.  Y 

Linear  Encoder  Input

 

7.  X  Linear  Encoder  Input

 

8.  Printer  Interface

 

9. 

Foot Switch connector

 

10. USB Connector

3

Layout of the Screen:

1

2

3

4

5

6

7

8

10

9

1. Unit
2

.

Measurement Mode

3

.

Coordinate System

4

.

Permenent Storage

5

.

Message Window

6

.

X Axis

7

.

Y Axis

8

.

Angular display 

9

.

Scroll Soft Key status

10

.

Soft Key

The setup can be accessed by pressing the     followed by key            .

 

The  user  will  be  prompted  for  password.  The  password  is  “98563”.   
Use [      ]or [      ] key to scroll within the menu.

Setup

24

Point Tolerance

:

Highlight  the  desired  point  feature  from  the  result  list.  Press                     
{             } key and then Press {             } key.

1

Enter the nominal values for X and Y co-ordinates. Also enter the
tolerance zone.
Tolerance  zone  is  the  radius  within  which  the  point  should  be 
located. If the point is located outside the zone, tolerance fails.
Press {          } key then enter

 

the

 

nominal

 

values

 

for

 

X

 

and

 

Y

 

co

-

ordinates

Also

 

enter

 

the

  + /- 

values

 

for

 

each

 

nominal

 

value

.

Limits allow the user to set upper and lower limits on a value.
For  example,  a  nominal  value  of  10.3250  might  have  an  upper 
limit  of  10.3260  and  a  lower  limit  of  10.3240.  For  example,  if  a 
part drawing gives a +/- tolerance of 0.005, simply enter 0.005 in 
the plus and minus fields.

After set press {            } key. The result of tolerance Pass or Fail is 
shown for the selected feature.

Use  the  same  way  to  set  Tolerance  for  line,  angle,  circle  and 
distance.

2

Maintenance

1

The  projector  is  a  kind  of  precision  optical  instrument.  Attention 
should  be  paid  to  its  maintenance  so  as  to  guarantee  the  precision 
and  lengthen  life  of  the  instrument.  The  instrument  should  be 
installed  in  a  clean  room,  the  temperature  of  which  should  be 
maintained at 20

° ±

 5

. The relative humidity of the room should 

not  exceed  60

  so  as  to  prevent  the  optical  parts  and  reflecting 

mirror from molding and to keep the high quality of the image.

Содержание ISP-A5000

Страница 1: ...www insize com ISP A5000 A5000E PROFILE PROJECTOR OPERATION MANUAL V0 MN ISP A5000 A5000E E MN ISP A5000 A5000E E...

Страница 2: ...tility Keys 8 Numeric Function Keys 9 Coordinate And Unit Switch Buttons 10 Axis Reset Keys Point Feature key Line Feature Key Circle feature key Distance Feature Key Angle Feature Key Skew Feature Ke...

Страница 3: ...tween two circles or lines 25 The surfaces of optical parts should be kept clean They should not be touched with hands If any dirt oil fleck or dust are found on the surfaces they should be removed wi...

Страница 4: ...hould be located If the point is located outside the zone tolerance fails Press key then enter the nominal values for X and Y co ordinates Also enter the values for each nominal value Limits allow the...

Страница 5: ...esults we must do the cross calibration process for every magnifications If we don t execute cross cal and if we probe a circle by cross hair and by edge detector in both cases we will get diameter eq...

Страница 6: ...his sets the gain of the amplifier The gain depends on the magnification and the light source Adjust this such that the Screen intensity on light region displays counts above 3000 counts to 4000 In ca...

Страница 7: ...to user defined segments The error in each segment is compensated with a single correction factor Each correction point is measured with respect to the starting point This starting point is usually se...

Страница 8: ...the next segment by pressing Next Calibration Edit Procedure When all the segments are complete press Enter key 20 Edge Detector For ISP A1000E Optical Edge detection can be used with DRO to increase...

Страница 9: ...nd measured X and Y coordinates from the standard during the setup process NLEC requires a repeatable machine zero defined on startup 3 The changes will get saved in the memory and are applied The use...

Страница 10: ...ual softkey Display value will show counts of respective rotation User will enter angle of rotation in seconds Now user can calculate calibration factor by pressing calc softkey After calibration pres...

Страница 11: ...t on the line and match cross hair on it Press key A Now in similar manner probe sufficient points to form a best fit line And then press key A Circle Measurement A circle can be measured by selecting...

Страница 12: ...al view of results will be displayed Use up down arrow keys to move next or previous results one by one or press key to recall particular result 9 Exit from result buffer view Visible only if result i...

Страница 13: ...eature measurement Also it automatically detects which geometrical shape has been probed This feature can be activated once or repeat 8 15 Skewing the job is also possible from constructed features Th...

Страница 14: ...Note As shown in the above figure the part must be placed within 45 degrees of the machine s co ordinate system Procedure for skewing the job Press key Following screen will be displayed Press key to...

Отзывы: