Calibration Procedure:
Keep the encoder position near to the desired reference on the
encoder.
P r e s s { } k e y i n “ M a c h i n e R e f . ” i n a x i s s e t t i n g s .
“Homing”Message is shown on the screen.
Pass the reference mark on the encoder. Here the DRO will reset
the axis on the reference mark.
Go to the machine reference position and press { }key.
Normally this is the start position of the segnments
Calibration procedure for Z (Angular) Axis
Manual Calibration of Z axis:
User can calibrate angular axis manually by pressing manual
softkey. Display value will show counts of respective rotation.
User will enter angle of rotation in seconds. Now user can
calculate calibration factor by pressing calc softkey.After
calibration press [Enter] to save the calibration factor. Press
[Cancel] if the user wishes to exit without saving.
1
Auto
Calibration
of
Z
axis
:
User can calibrate the angular axis automatically by selecting the
{ }key. The DRO waits for two complete rotations of the
encoder and then calculates the calibration factor by calculating
the counts between the two reference marks on the encoder. To
calibrate the Z axis using auto mode requires only one condition
that it must have only one reference mark for each rotation.
2