Normal heart rhythm is detected:
If a normal heart rhythm is detected, the defibrillator will start an internal
discharging.
Discharging Procedure
When the charging procedure of the defibrillator is finished, the defibrillator
will enter the discharging procedure and give voice prompts to the operator.
After delivering a shock, the defibrillator will initiate the CPR procedure and
will provide CPR guidance to the user.
Cardiopulmonary Resuscitation (CPR) Procedure
SALI-D will enter cardiopulmonary resuscitation procedure when the condition
"No shockable heart rhythm" is detected.
During the heart rhythm analyzing period, if the heart rhythm is not considered
to be a shockable rhythm by the defibrillator, a cardiopulmonary resuscitation
procedure (CPR) will start.
During the period of charging, if the defibrillators detects that the shockable
rhythm has changed to a normal one, the defibrillator will stop the current
procedure and switch to CPR procedure.
Warning:
During the discharging procedure the following
instruction will appear "Don't touch patient, rhythm analysis"
during that time the patient should not be touched, the
operator should clear any bystanders near the patient before
pressing the red shock button.
Note:
During the cardiopulmonary resuscitation period, the AED will
ignore the heart rhythm analyzed result and the improper placement of the
AED pad connection.
If the patient is not breathing and has no pulse, a cardiopulmonary
resuscitation should be performed on the patient immediately.
Note:
The CPR time is fixed at 120 seconds.
At the end of a CPR cycle, the AED will give a prompt to indicate that the
operator must stop CPR and not touch the patient so that the AED can re-
analyze the heart rhythm and determine if a shockable rhythm exists.
Содержание SALI D
Страница 1: ...Page 1 of 48 EMPOWERING EMERGENCY CARE INV IFU 2001 Rev 09 Last update September 2021...
Страница 24: ...Page 24 of 48 8 Battery and SIM compartment 9 O2 Cylinder compartment 10 Docking Station charging points 8 9 10...
Страница 37: ...Page 37 of 48 Hand placement Compression rhythm Stop Responsive Patient Place Mask on face Place ECG electrodes...
Страница 38: ...Page 38 of 48 Place SpO2 sensor...
Страница 46: ...Page 46 of 48 Fax 972 9 965 64 79 E mail Info Inovytec com Web Site http www Inovytec com...