Rhythm Recognition Performance
The AED algorithm exceeds the requirements of ANSI/AAMI DF39-1993,
section 3.3.18 and the sensitivity and specificity levels recommended by the
AHA - Automatic External Defibrillators for Public Access Use:
Recommendations for Specifying and Reporting Arrhythmia Analysis Algorithm
Performance.
The test database includes shockable rhythms consisting of ventricular
fibrillation rhythms (>150uV) and wide-complex ventricular tachycardia at a
rate greater than 160 BPM. Non-shockable rhythms include various sinus
rhythms including supraventricular tachycardia, atrial fibrillation, atrial flutter,
sinus rhythm with PVC's, asystole, pacemaker rhythms, and ventricular
tachycardia with a rate less than 160 BPM and/or narrow complexes.
Rhythms
Test Sample
Size
Performance Goal Conclusion
Shockable
rhythm: VF
1067
>
90% sensitivity
Meets the AAMI
DF39 requirement
and AHA
recommendation
Shockable
rhythm: VT
22
>
75% sensitivity
Non-Shockable
rhythm: NSR
4000
>
99% sensitivity
(AHA)
Non-Shockable
rhythm: Asystole
179
>
95% sensitivity
Non-Shockable
rhythm: All other
rhythms
25732
>
95% sensitivity
Содержание SALI D
Страница 1: ...Page 1 of 48 EMPOWERING EMERGENCY CARE INV IFU 2001 Rev 09 Last update September 2021...
Страница 24: ...Page 24 of 48 8 Battery and SIM compartment 9 O2 Cylinder compartment 10 Docking Station charging points 8 9 10...
Страница 37: ...Page 37 of 48 Hand placement Compression rhythm Stop Responsive Patient Place Mask on face Place ECG electrodes...
Страница 38: ...Page 38 of 48 Place SpO2 sensor...
Страница 46: ...Page 46 of 48 Fax 972 9 965 64 79 E mail Info Inovytec com Web Site http www Inovytec com...