3.3
LIMITATIONS OF USE
Clinical situations potentially affecting accuracy or performance:
•
Controlling the flow in the presence of difficult airways, such as severe
lung blockage and asymmetric air entrance to the lung.
•
Low compliance of the airways.
•
Laryngospasm and or water in the upper and lower air way as well as
edema may interfere with the device ability to monitor the patient
breath rate.
•
Hypothermia can cause vasoconstriction which will interfere with the
SPO2 measurements.
•
Drug abuse may cause vasoconstriction which will interfere with the
SPO2 measurements.
•
When treating COPD patients, the targeted SpO
2
measurement is 88-
92% this may affect the oxygen administration as the SALI-D algorithm
stipulate an increase of O
2
flow to 10 LPM as such, in cases where the
saturation levels are below 90, the increase in O
2
flow may potentially
decrease respiration rate.