Warnings
Do not use alcohol, iodine or other skin preparations.
Before defibrillating, disconnect other electrical equipment which
has no DEFIBRILLATION-PROOF applied parts from the patient.
DO NOT allow defibrillation pads to touch each other, or to touch
other electrodes, lead wires, dressings, transdermal patches, etc.
During defibrillation, the operator and all other people must stand
clear of the patient, bed, and all conductive surfaces in contact with
the patient.
The Defibrillator contains an automatic disarm of the stored energy.
If the operator has not delivered the energy to a patient or a test
load, an internal timer will disarm the stored energy. This stored
electrical energy can potentially cause death or injury if discharged
improperly. Follow all instructions in this User’s Manual.
DO NOT move patient. Handling or transporting the patient during
ECG analysis can cause incorrect or delayed diagnosis
Cardiac pacemakers may affect rhythm analysis. Patient pacemakers
may reduce the sensitivity of device analysis and errors in detecting
shockable rhythms.
Radio frequency (RF) interference. Do not operate the defibrillator in
conjunction with electrocautery or diathermy equipment. Any
equipment that emits strong radio frequency signals can cause
electrical interference and distort the analysis signal to cause
inaccurate interpretation of rhythm.
Possible explosion and fire hazard if used in areas where flammable
agents are used.
DO NOT immerse or clean pads with alcohol or solvents
DO NOT perform chest compressions (CPR) through pads. These
actions may damage the electrode pads and cause the Jump Start to
function improperly.
Содержание SALI D
Страница 1: ...Page 1 of 48 EMPOWERING EMERGENCY CARE INV IFU 2001 Rev 09 Last update September 2021...
Страница 24: ...Page 24 of 48 8 Battery and SIM compartment 9 O2 Cylinder compartment 10 Docking Station charging points 8 9 10...
Страница 37: ...Page 37 of 48 Hand placement Compression rhythm Stop Responsive Patient Place Mask on face Place ECG electrodes...
Страница 38: ...Page 38 of 48 Place SpO2 sensor...
Страница 46: ...Page 46 of 48 Fax 972 9 965 64 79 E mail Info Inovytec com Web Site http www Inovytec com...