28
Surge
IEC 61000-4-5
± 1 kV line(s)
to line(s)
± 2 kV line(s)
to earth
Not
applicable
Not applicable.
This system does not
connect to mains power or
to earth ground.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip
in UT) for 0,5
cycle
40 % U
T
(60 % dip in
UT) for 5 cycles
70 % U
T
(30 % dip in
UT) for 25
cycles
<5 % U
T
(>95 % dip in
U
T
) for 5 s
Not
applicable
Not applicable.
This system does not
connect to mains power.
Power frequency (50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE: UT is the A.C. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration – electromagnetic immunity
The TAV® system is intended for use in the electromagnetic environment specified below.
The user of this system should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compli-
ance level
Electromagnetic
environment -
guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80
MHz
Not
applicable
Not applicable.
This system does not
connect to mains power or
to earth ground.
Appendix A: EMC Information