17
*
Connect the nasal pillow interface and gas source to the TAV® device.
*
Turn the gas source and TAV® device on.
*
Place the TAV® device in the Constant flow mode.
*
Turn the flow setting up to 5 for at least 10 seconds until any trapped water has
been cleared from the tubing.
º
When finished, make sure your nasal pillow interface is clean and free of debris.
If not, clean again.
Monthly (or as directed by your health-care provider):
•
Replace your nasal pillow interface with a fresh one supplied by your medical supplier.
WARNING:
Do not subject the Nasal Pillows Interface to heat sterilization, hot water
pasteurization, autoclaving, radiation sterilization, ethylene oxide gas sterilization, or attempt
to clean it in a dishwasher or microwave oven. Doing any of these may damage the nasal pillow
interface and impair oxygen delivery.
General Troubleshooting
Problem
Possible Causes
What To Do
Device
won’t turn
on
No battery installed
Install battery as shown on page 6.
Battery installed backwards
Check to make sure spring contact
is against “-“ (flat) side of battery.
Remove and flip battery if
necessary.
Power button was not held down
long enough
Hold power button down for one
full second until lights flash and a
beep is heard.
Battery not making good contact
with terminals
Remove and reinsert battery as
shown on page 6.
Dead battery
Remove battery and replace it with
a new one as shown on page 6.
Device malfunction
•
Open the bypass by pulling
out the stem (page 14). The
TAV® will deliver a constant
flow at 2 LPM.
•
Contact Customer Support
at 888-599-1105 as soon as
possible.
Device
won’t turn
off
Power button was not held down
long enough
Hold power button down for 3 full
seconds until lights turn off and
the device beeps twice.