7
Connecting to Gas Sources with DISS Fittings
Note: Do not connect the TAV® system to
the standard oxygen barb on the regulator
(shown below). Always use the provided
regulator adapter when connecting to
pressure sources greater than 28psig.
1. Attach one end of the supply tubing to
the connector on the bottom right of
the device.
2. Attach the other end of the supply
tubing to the Regulator Adapter. Make
sure that both ends of the supply tubing
are firmly pressed onto each connector.
3. Always ensure the blue Regulator
Adapter is connected to the TAV® device before connecting to the gas source.
The TAV® device is connected to standard DISS fittings using the Regulator Adapter accessory.
The Regulator Adapter must be used when connecting the TAV® device to hospital wall oxygen.
Attach the TAV® Regulator Adapter to the threaded DISS outlet of the gas source. Hand tighten
the connector until it is snug. Always ensure the Regulator Adapter is connected to the TAV®
device before connecting to the gas source.
Note: To ensure proper operation, the TAV® Regulator Adapter must be used when connecting
to DISS connectors or to hospital wall outlets that output 40 to 87 psig with greater than 20 LPM
capability.
Gas sources must be equipped with a standard threaded male DISS connector in order to
connect the TAV® Regulator Adapter. DISS = Diameter Index Safety System.