Alarms and Low Battery Indicator
The TAV® device alerts you if the flow output from the device is too low or too high compared
to your selected settings. It also provides a lighted indicator when the battery needs to be
replaced.
Low Battery Indicator: The
yellow low battery light will
turn on when the battery has
about 2 days of life remaining.
Replace the battery at your
soonest convenience.
Breath Alarm: This alarm
indicates that the user is not
triggering the device or the
breath rate is too fast or too
slow for the current setting.
You will see the yellow “breath
alarm” light flash and hear a
series of beeps.
Flow Alarm: This alarm
indicates that the flow output
is not the same as the set
flow. You will see the red “flow
alarm” light flash and hear a
series of beeps
What to do:
1. Check the patient
2. Check nasal pillow interface fit
3. Check tubing for kinks and leaks.
4. Check that all tubing connections are
securely attached.
5. Turn the device OFF, and then back ON.
6. If the problem persists, turn off the
device and contact customer support
for assistance. Use Bypass Valve if
needed.
What to do:
1. Check tubing for kinks and leaks.
2. Check that all tubing connections are
securely attached.
3. Check gas source
4. If using a cylinder regulator, make sure
the Regulator Adaptor is being used.
5. Turn the device OFF, and then back ON.
6. If the problem persists, turn off the
device and contact customer support
for assistance. Use Bypass Valve if
needed.
To silence an alarm: Briefly press the button to silence the alarm for 1 minute. If the
problem persists, the alarm will sound again after 1 minute.
Note: The alarm light will stay on until the problem is corrected.
Warning: A brief press of the alarm button will silence the alarm - the alarm light will stay on.
Holding down the button for longer than 3 seconds will turn off the TAV® device.