![Innomed Medical HeartScreen112 Clinic Скачать руководство пользователя страница 56](http://html.mh-extra.com/html/innomed-medical/heartscreen112-clinic/heartscreen112-clinic_user-manual_2062871056.webp)
HS112 C-1 User Manual
56/73
R-6730-EN02-E
13.4. Language
Setting the device language and the date display format:
1:
The display language of the menu can be set here. (English)
The notifications on the display and printout will appear in the language selected here.
By changing the device language, the date format will change accordingly, as different
languages have different layouts.
2:
Date:
The date format of the respective language can be changed, overwritten. The
three possible date formats are:
year, month, day;
month, day, year (default setting);
day, month, year.
13.5. Filters
1:
Network:
Setting the frequency of the mains noise filter. The possible choices are 50
Hz and 60 Hz (Europe, America). (50 Hz)
2:
Baseline:
Enabling the baseline filter. (On)
This option is activated every time the device is turned on.
This setting does not correspond to that of the display baseline filter.
13.6. Others
1:
QRS volume:
0; 1; 2; 3; (0) –– 0 is mute
2:
Key sound:
0; 1; 2; 3; (1) –– 0 is mute
3:
Backlight
: 0; 1; 2; 3; (2) The backlight of
the display. Set to 2 every time the device
is turned on. In battery mode, the battery
lasts longer if this value is lower.
4:
Colors:
0; 1; 2; 3; (0) –– The colour of the
display. When turning the device on, the
last setting is used.
5:
Turn off:
Enabling or disabling automatic
device shutdown. If this setting is enabled, the device shuts down after 5
minutes if no buttons are pressed and there is no ECG signal or an ongoing PC
connection. (5 minutes)
6:
Leads:
Lead groups may be standard or Cabrera. (standard)
7:
HR warning:
Enabling or disabling the HR warning (the limit values can be
set in the
Patient ID
menu). (Off) See section 8.2.
8:
Memory clear:
A complete memory reset. All stored records are deleted, the
settings are returned to their default values. The reset can be performed by
pressing the LEAD button.