11.2. Deleting a record
The selected record can be deleted by pressing the
DEL (AUTO)
button. The device will
request a confirmation by asking: “Are you sure you want to delete?”
“
No: AUTO Yes: LEAD
” will appear at the bottom of the display. If the
LEAD
button is
pressed, the record is deleted, and the catalogue without the deleted record is shown on the
display. If the
AUTO
button is pressed, the record remains in the catalogue.
It is not necessary to delete records manually as the device deletes the oldest record
automatically in case the memory is full.
11.3. Printing all record data
Pressing the
button in the main menu of the catalogue when the list of records is
displayed prints out all stored elements of the selected record (indicated by the red record
number). The elements are printed in the following order — if available:
Diagnosis program
Header
Stored ECG record of the 12 channels
Rhythm curve (with a 5 mm/s sweep speed)
Average curves (with a fixed 25 mm/s sweep speed)
Analysis data table
Diagnosis (Stored only if a diagnosis was successfully created.)
Store program
Header
Stored ECG record of the 12 channels
Rhythm curve (with a 5 mm/s sweep speed)
Store program with arrhythmia analysis
Header
Stored ECG record of the 12 channels
Rhythm curve
Arrhythmia summary data
Arrhythmia diagrams
Printing can be stopped by pressing the
button one more time.