DOCUMENT NO. 920012 REV 2 EFFECTIVE 03-14-18 INFRASCAN CONFIDENTIAL PAGE 30 OF 35
Attenuator Shield Test
Attenuator Shields
(Part Number 000036)
are available
for purchase for service personnel to do periodic testing
of the Infrascanner. The Attenuator Shields are
calibrated by the factory and can be used on a periodic
basis (e.g. annually) to check that the Infrascanner is
functioning properly. To verify that the Infrascanner
functioning properly, place the Attenuator Shield on the
Infrascanner. Navigate to the Measurement Screen and
take two frontal measurements. Record the Attenuator
Shield nominal attenuation and the data file number.
Remove the Attenuator Shield.
Place the Infrascanner in a Cradle connected to a PC
running Windows 7 or later. Turn on the Infrascanner by
depressing the Cradle “ON” button. When the PC recognizes
the Infrascanner with the Windows Mobile Device Center, download the data file created
above from the HS Data folder on the SD Storage Card, and open it. The file name
includes a measurement number, date, and time.
Record the actual OD measured. It must be within ±0.3 of the nominal attenuation marked
on the Attenuator Shield.
Electromagnetic Interference
The essential performance is pulsing the laser to perform a measurement.
Guidance and manufacturer’s declaration – electromagnetic emissions
The Model 2000 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model 2000
should assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The Model 2000 must emit electromagnetic energy in order to perform its intended
function. Nearby electronic equipment may be affected.
RF emissions
CISPR 11
Class A
The Model 2000 is suitable for use in all establishments other than domestic
establishments and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 60601-3-2
Not
Applicable
Voltage fluctuations /flicker
emissions IEC60601-3-2
Not
Applicable
Guidance and manufacturer’s declaration – electromagnetic immunity
The Model 2000 is intended for use in the electromagnetic environment specified below. The customer or the user of the Model
2000 should assure that it is used in such an environment.
Immunity test
Test Level
Compliance
Electromagnetic environment - guidance
ESD
IEC 61000-4-2
+6KV Contact
+8KV Air
+6KV Contact
+8KV Air
Passed: Criteria C
Floors should be wood, concrete or ceramic tile. If
floors are synthetic the relative humidity should be a
least 30%.
EFT
Not Applicable
Not Applicable
Not Applicable
Figure 5-1 Attenuator
Shield