DOCUMENT NO. 920012 REV 2 EFFECTIVE 03-14-18 INFRASCAN CONFIDENTIAL PAGE 33 OF 35
6.0
INFRASCANNER MODEL 2000 TECHNICAL SPECIFICATIONS
1.
Physical Specifications:
1.1.
Dimensions with Disposable: 9.0H x 3.4W x 2.4D inch (23H x 8.5W x 6.1D cm)
1.2.
Dimensions without the Disposable Fiber Optic Shield: 9.0H x 3.4W x 1.8D inch
(23H x 8.5W x 4.6D cm)
1.3.
Weight of Infrascanner: 14oz (400g)
1.4.
Number of detectors: 1
1.5.
Number of light sources: 1
1.6.
Light Source - Detector Separation: 4 cm
2.
Functional Specifications
2.1.
Power 4.8V, Rechargeable NiMH Battery pack (1AH) or 4 AA disposable
batteries
2.2.
Standby time 2 weeks.
2.3.
Operating time on new fully charged batteries: >60 minutes
2.4.
Operating time on new AA batteries: >90 minutes
2.5.
Hematoma indication:
ΔOD > 0.2 +/-0.02
3. Laser Diode Specifications:
3.1.
Wavelength: 808nm (±3nm)
3.2.
Maximum peak pulsed power: 100 milliwatt
3.3.
Safety: Class 1 Laser Product
4. Cradle Power Adapter Specification:
GlobTek, Inc. 186 Veterans Dr. Northvale, NJ 07647, 201.784.1000. Model Number:
WR9QA1200LCP-N-MED (R). Input voltage: 100-240 volt AC. Input current: < 0.5 Amps.
Input frequency: 50-60 Hz. Output: 5 VDC @ 1.2 amps.
5. Environmental:
Operational
Storage
Temperature
-10°C to 50°C
-40°C to 70°C
Relative Humidity (non-cond)
0% to 90%
0% to 90%
Atmospheric Pressure
70-106 kPa
50-106 kPa
Complies with IEC 60601-1 3
rd