DOCUMENT NO. 920012 REV 2 EFFECTIVE 03-14-18 INFRASCAN CONFIDENTIAL PAGE 18 OF 35
for novice users. Once the user is more experienced with the use of the
Infrascanner, Independent Mode should be selected.
Figure 3-8: Guided Mode Measurement Windows
5.
Click on the “Next” button to start patient measurement. After the appearance of
the Measurement Screen (Figure 3-9)
verify that the “OK, Waiting probe” message
appears on the bottom left of the screen.
Verify that “Coupled Mode” is indicated on
the top left of the screen.
If not, contact the factory. If a
“communications error”
message is indicated
instead of the “OK, Waiting probe” message, remove the
Disposable Fiber Optic Shield and put it back on to
reset the Infrascanner. Re-initiate the data collection.
6.
The Infrascanner is ready to begin measuring. Take
the Infrascanner, and start the head scan, alternating
between left and right positions according to the head
scanning sequence in Figure 2-4. The blue square on
the screen in Figure 3-9 indicates the current
measurement location.
7.
In each location wiggle the light guides so they will be in
clear contact with the scalp. Ascertain that no hair is
between the light guides and the scalp. After
establishing firm contact press and release one of the
two
“Measure” Buttons on the rear of the Infrascanner.
Either one of the buttons can be used, depending on
which is more convenient. The measurement begins
after the button is released and an audible beep is
sounded.
Use your free hand to support patient’s head,
by placing it on the contra-lateral side of the
measurement.
Figure 3-9: Measurement
Screen